FDA Adverse Event Injury Summary report: N

OT ULTRALINK METER

MDR report key: 3171431 · Received June 15, 2013

Report

Report Number
3008382007-2013-17325
Event Type
Injury
Date Received
June 15, 2013
Date of Event
June 3, 2013
Report Date
June 3, 2013
Manufacturer
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
Product Code
NBW
PMA / PMN Number
K073231
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP #1 (07/02/2013)-DEVICE EVALUATION: THE METER AND TEST STRIPS INVOLVED WITH THIS COMPLAINT HAVE BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS ON 6/25/2013 AND 6/27/2013, RESPECTIVELY, WITH THE FOLLOWING FINDINGS: THE METER PASSED ALL TESTING WITH NO FAULTS FOUND; THE COMPLAINT WAS NOT CONFIRMED. THE TEST STRIPS WERE ALSO TESTED AND THE PRIMARY COMPLAINT WAS NOT CONFIRMED. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.

Additional Manufacturer Narrative · 1

LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT.

Description of Event or Problem · 1

ON (B)(6) 2013, THE LAY USER/PATIENT CONTACTED LIFESCAN (LFS) ALLEGING THAT HIS ONETOUCH ULTRALINK METER WOULD NOT POWER ON. THE COMPLAINT WAS CLASSIFIED BASED ON THE CUSTOMER CARE ADVOCATE (CCA) DOCUMENTATION. THE PATIENT REPORTED THAT THE ALLEGED POWER ISSUE BEGAN ON (B)(6) 2013, AT 7PM. THE PATIENT IS ON INSULIN PUMP THERAPY. THE PATIENT DENIED TAKING ANY ACTION REGARDING HIS USUAL DIABETES MANAGEMENT REGIMEN DUE TO THE METER ISSUE. HOWEVER, THE PATIENT CLAIMED FEELING SHAKY 12 HOURS LATER. THE PATIENT DENIED RECEIVING MEDICAL TREATMENT. AT THE TIME OF TROUBLESHOOTING, THE CCA NOTED THAT THE SUBJECT METER WOULD POWER ON MANUALLY (WHEN POWER BUTTON PRESSED), BUT WOULD NOT POWER ON WHEN A TEST STRIP WAS INSERTED. THE CCA NOTED THAT THE PATIENT WAS USING THE CORRECT TEST STRIPS AND INSERTING THE STRIP CORRECTLY. REPLACEMENT PRODUCTS WERE SENT TO THE PATIENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE PATIENT REPORTEDLY DEVELOPED SIGNS/SYMPTOMS SUGGESTIVE OF A SERIOUS INJURY AFTER THE ALLEGED METER POWER ISSUE STARTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
272694 OT ULTRALINK METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL 3361987

Patients

Seq Age Sex Outcome Treatment
1 Life Threatening