SCREWDRIVER
Report
- Report Number
- 1030489-2013-02314
- Event Type
- Malfunction
- Date Received
- June 14, 2013
- Date of Event
- May 17, 2013
- Report Date
- July 9, 2013
- Manufacturer
- MEDTRONIC SOFAMOR DANEK USA, INC
- Product Code
- HXX
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(6). (B)(4). THE DEVICE HAS BEEN RETURNED TO THE MANUFACTURER FOR EVALUATION. ANALYSIS RESULTS ARE NOT AVAILABLE AT THE TIME OF THIS REPORT. A FOLLOW-UP REPORT WILL BE SENT WHEN THE ANALYSIS IS COMPLETE.
THE DEVICE HAS BEEN RETURNED TO THE MANUFACTURER FOR EVALUATION. VISUALLY CONFIRMED APPROXIMATELY ~3MM OF THE INSTRUMENT TIP HAS BEEN BROKEN OFF, CONSISTENT WITH INTERFACE DURING USAGE. DIMENSIONAL INSPECTION OF THE INNER SHAFT DIAMETER CONFIRMED CONFORMANCE TO PRINT SPECIFICATION. FRACTURE SURFACE ANALYSIS REVEALS A FAIRLY FLAT FRACTURE SURFACE AND CIRCULAR MATERIAL MOVEMENT, CONSISTENT WITH TORSIONAL OVERLOAD, CONSISTENT WITH TORSIONAL OVERLOAD CONDITION. THE ABOVE FINDINGS ARE CONSISTENT WITH TORSIONAL OVERLOAD.
IT WAS REPORTED THAT THE PATIENT¿S INITIAL DIAGNOSIS WAS A TRAUMATIC INJURY WITH FIXATION LEVEL 2-ABOVE AND 2-BELOW ON L3. IT WAS REPORTED THAT THE TIP OF THE DRIVER BROKE DURING FINAL TIGHTENING. NO PATIENT COMPLICATIONS WERE REPORTED.