FDA Adverse Event Malfunction Summary report: N

COULTER® LH 780 HEMATOLOGY ANALYZER

MDR report key: 3171376 · Received June 14, 2013

Report

Report Number
1061932-2013-01163
Event Type
Malfunction
Date Received
June 14, 2013
Date of Event
May 16, 2013
Report Date
May 16, 2013
Manufacturer
BECKMAN COULTER
Product Code
GKZ
PMA / PMN Number
K061616
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE FIELD SERVICE ENGINEER (FSE) DISCOVERED A PIN HOLE IN THE RED STRIPE TUBING FROM THE DIFFERENTIAL HEATER TO THE SHEAR VALVE; THE FLUID WAS DIFFERENTIAL LYSE REAGENT. THE FSE REPLACED THE AFFECTED TUBING AND RESOLVED THE ISSUE. THE FSE PERFORMED SYSTEM SHUTDOWN AND STARTUP, LATRON CONTROL, 5C AND RETIC CONTROLS, AND A PRECISION TEST; ALL RESULTS WERE WITHIN SPECIFICATIONS. SERVICE ACTIVITY PERFORMED WAS VERIFIED TO MEET THE SPECIFIED REQUIREMENTS PER ESTABLISHED PROCEDURES. THE INSTRUMENT CONFORMED TO THE MANUFACTURER'S PUBLISHED PERFORMANCE SPECIFICATIONS AND WAS RETURNED TO NORMAL OPERATION. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED APPROXIMATELY TWO MILLILITERS OF DILUENT LEAKED OUTSIDE THE INSTRUMENT, UNDER THE SHEAR VALVE, INVOLVING THE COULTER LH 780 HEMATOLOGY ANALYZER. THE CUSTOMER WAS WEARING PROPER PERSONAL PROTECTIVE EQUIPMENT (PPE) CONSISTING OF GLOVES, A LABORATORY COAT, AND EYE PROTECTION AND DID NOT HAVE DIRECT CONTACT WITH THE FLUID. THERE WAS NO OPERATOR INJURY OR ADVERSE EFFECT ASSOCIATED WITH THIS EVENT. PATIENT RESULTS WERE NOT IMPACTED. THERE WAS NO PATIENT CONSEQUENCE. THE FACILITY HAS AN EXPOSURE CONTROL OR RISK MANAGEMENT PLAN IN PLACE. A BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO ASSESS THE INSTRUMENT.