FDA Adverse Event Malfunction Summary report: N

COULTER® LH 780 HEMATOLOGY ANALYZER

MDR report key: 3171375 · Received June 14, 2013

Report

Report Number
1061932-2013-01161
Event Type
Malfunction
Date Received
June 14, 2013
Date of Event
May 16, 2013
Report Date
May 16, 2013
Manufacturer
BECKMAN COULTER
Product Code
GKZ
PMA / PMN Number
K061616
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE FIELD SERVICE ENGINEER (FSE) NOTED FLUID LEAKED AT PINCH VALVE VL82 (COMPLETE BLOOD COUNT (CBC) LYSE LINE). THE FSE REPLACED THE TUBING AND RESOLVED THE FLUID LEAK ISSUE. THE FSE COMPLETED SYSTEM STARTUP, LATRON CONTROL, QUALITY CONTROL (QC) AND REPRODUCIBILITY TEST; ALL RESULTS WERE WITHIN SPECIFICATIONS. THE INSTRUMENT CONFORMED TO THE MANUFACTURER'S PUBLISHED PERFORMANCE SPECIFICATIONS AND WAS RETURNED TO NORMAL OPERATION. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED APPROXIMATELY FIFTEEN (15) MILLILITERS OF CLEAR FLUID LEAKED ONTO THE COUNTER INVOLVING THE COULTER LH 780 HEMATOLOGY ANALYZER. THE CUSTOMER STATED THE LEAK ORIGINATED FROM PINCH VALVE TUBING VL82 DUE TO A CUT IN THE TUBING. THE CUSTOMER WAS WEARING PROPER PERSONAL PROTECTIVE EQUIPMENT (PPE) CONSISTING OF GLOVES, A LABORATORY COAT, AND EYE PROTECTION AND DID NOT HAVE DIRECT CONTACT WITH THE FLUID. THERE WAS NO OPERATOR INJURY OR ADVERSE EFFECT ASSOCIATED WITH THIS EVENT. PATIENT RESULTS WERE NOT IMPACTED. THERE WAS NO PATIENT CONSEQUENCE. THE FACILITY HAS AN EXPOSURE CONTROL OR RISK MANAGEMENT PLAN IN PLACE. A BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO ASSESS THE INSTRUMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
271970 COULTER® LH 780 HEMATOLOGY ANALYZER COUNTER, DIFFERENTIAL CELL GKZ BECKMAN COULTER NA NA