SIMS GRASEBY
Report
- Report Number
- 1120886-2001-00001
- Event Type
- Other
- Date Received
- February 22, 2001
- Date of Event
- January 16, 2001
- Report Date
- February 22, 2001
- Manufacturer
- MARCAL MEDICAL/IMPORTER-DISTRIBUTOR, SIMS GRASEBY LTD.
- Product Code
- FRN
- Report Source
- Distributor report
- Reporter Location
- PA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
PUMP INFUSED A CARDIAC STRESS AGENT 30 SECONDS TOO FAST DURING A CARDIAC STRESS TEST. MARCAL'S BIOMEDICAL SERVICE TECH TROUBLE SHOT OVER THE PHONE WITH HOSPITAL BIOMEDICAL SERVICE TECH, AND DETERMINED THAT THERE WAS NO DEATH OR INJURY TO THE PT INVOLVED. HE ALSO DETERMINED THAT THE CUSTOMER WAS USING A 30CC SYRINGE. HOWEVER, THE SYRINGE SIZE SENSOR DISPLAY ON THE PUMP WAS INDICATING THAT A 20CC SYRINGE WAS BEING USED. THE PUMP WOULD INDEED INFUSE INCORRECTLY IF THE SYRINGE SIZE DISPLAY READ DIFFERENTLY FROM THE ACTUAL SIZE OF THE SYRINGE BEING USED. THE OPERATIONS MANUAL CLEARLY PROHIBITS USING THE PUMP IF THE SYRINGE SIZE SENSOR DISPLAY READING DOES NOT AGREE WITH THE SYRINGE SIZE BEING USED. THIS WAS AN OBVIOUS OPERATOR ERROR AND COULD HAVE BEEN PREVENTED IF THE USER HAD READ THE DISPLAY BEFORE USING THE PUMP. CO REQUESTED THE PUMP BE RETURNED TO MARCAL MEDICAL FOR REPAIR AND CALIBRATION. AN INCIDENT REPORT WAS FILLED OUT TO BE FOLLOWED UP ON WHEN INVESTIGATION COMPLETED. INVESTIGATION IN PROGRESS REPORT TO FOLLOW.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 6883 | SIMS GRASEBY | SYRINGE PUMP | FRN | MARCAL MEDICAL/IMPORTER-DISTRIBUTOR, SIMS GRASEBY LTD. | 3400 | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * | Other |