FDA Adverse Event
Injury
Summary report: N
SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
MDR report key: 3171232
·
Received June 14, 2013
Report
- Report Number
- 1416980-2013-15449
- Event Type
- Injury
- Date Received
- June 14, 2013
- Date of Event
- May 15, 2013
- Report Date
- May 22, 2013
- Manufacturer
- BAXTER HEALTHCARE
- Product Code
- KDJ
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE CAUSE OF THIS PERITONITIS WAS USE ERROR. PER BAXTER LABELING, USERS ARE INSTRUCTED TO USE ASEPTIC TECHNIQUE WHEN PERFORMING PERITONEAL DIALYSIS THERAPY. A FORMAL REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. IF THERE IS ANY FURTHER RELEVANT INFORMATION FROM THAT REVIEW, A SUPPLEMENTAL MEDWATCH WILL BE FILED.
Description of Event or Problem · 1
THIS IS A REPORT OF A HOME PATIENT (HP) WAS DIAGNOSED WITH PERITONITIS. THE CAUSE OF PERITONITIS WAS BREAK IN ASEPTIC TECHNIQUE, FURTHER DESCRIBED AS MISTAKE TOUCH CONTAMINATION. THE HP WAS REPORTED TO BE RECOVERING. ADDITIONAL INFORMATION WAS REQUESTED BUT NOT PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 271826 | SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE | KDJ | BAXTER HEALTHCARE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 46 YR | Other | DIANEAL, AND HOME CHOICE |