FDA Adverse Event Malfunction Summary report: N

PRECISION®

MDR report key: 3171158 · Received June 14, 2013

Report

Report Number
3006630150-2013-01225
Event Type
Malfunction
Date Received
June 14, 2013
Date of Event
May 20, 2013
Report Date
May 20, 2013
Manufacturer
BOSTON SCIENTIFIC NEUROMODULATION
Product Code
LGW
PMA / PMN Number
030017
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL SUSPECT MEDICAL DEVICE COMPONENTS INVOLVED IN THE EVENT: MODEL #: SC-2218-70, SERIAL #: (B)(4), DESCRIPTION: LINEAR ST LEAD, 70CM THE EXPLANTED DEVICES WERE NOT RETURNED TO BSN AS THEY WERE DISCARDED BY THE MEDICAL FACILITY. IT IS INDICATED THAT THE DEVICE WILL NOT BE RETURNED FOR EVALUATION; THEREFORE, A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. A REVIEW OF THE DEVICE HISTORY RECORDS WILL BE CONDUCTED. IF THERE IS ANY FURTHER RELEVANT INFORMATION FROM THAT REVIEW, A SUPPLEMENTAL MED WATCH WILL BE FILED.

Description of Event or Problem · 1

A REPORT WAS RECEIVED THAT THE PATIENT HAD INADEQUATE STIMULATION. IT WAS FOUND OUT THAT FOUR CONTACTS WERE OUT. THE PATIENT UNDERWENT REVISION WHEREIN THE LEAD WAS REPLACED. THE PATIENT WAS DOING WELL FOLLOWING THE REVISION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
271243 PRECISION® SPINAL CORD STIMULATOR LGW BOSTON SCIENTIFIC NEUROMODULATION SC-2218-70 NA

Patients

Seq Age Sex Outcome Treatment
1 62 YR