FDA Adverse Event Malfunction Summary report: N

STREAMLAB® ANALYTICAL WORKCELL

MDR report key: 3171149 · Received June 14, 2013

Report

Report Number
1226181-2013-00266
Event Type
Malfunction
Date Received
June 14, 2013
Date of Event
May 18, 2013
Report Date
May 20, 2013
Manufacturer
SIEMENS HEALTHCARE DIAGNOSTICS
Product Code
LXG
PMA / PMN Number
EXEMPT
Removal / Correction Number
1226181-10/17/2012-002-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A SIEMENS HEALTHCARE DIAGNOSTICS TECHNICAL SERVICE CONSULTANT (TSC) WAS CONTACTED BY THE CUSTOMER. THE TSC DETERMINED THAT THE CAUSE OF THE OPERATOR FALLING, WAS DUE TO A PROTRUSION OF A PIN BELOW THE FRAME OF THE INPUT/OUTPUT MODULE. AN URGENT CUSTOMER NOTIFICATION, (B)(4); OCTOBER 2012, WAS PROVIDED TO ALL STREAMLAB AUTOMATION SOLUTIONS CUSTOMERS TO INFORM THEM THAT SIEMENS HEALTHCARE DIAGNOSTICS HAS RECEIVED REPORTS FROM CUSTOMERS THAT THE SMALL PINS AT THE BOTTOM OF THE CENTER DOOR PANEL OF THE INPUT/OUTPUT MODULE, WHICH PROTRUDE APPROXIMATELY 5/8 INCH (16 MM), HAVE CONTRIBUTED TO TWO TRIP AND FALL INCIDENTS. CUSTOMERS ARE ADVISED THAT WHEN WORKING IN THE AREA OF THE INPUT/OUTPUT MODULE TO BE AWARE OF THE TWO PINS THAT PROTRUDE AND TO USE CAUTION. THEY ARE ASKED TO NOTIFY ALL STREAMLAB OPERATORS IN THE LABORATORY TO THIS POTENTIAL SAFETY ISSUE AND TO POST THE NOTICE ON OR NEAR THE SYSTEM.

Description of Event or Problem · 1

AN OPERATOR WAS PERFORMING STANDARD OPERATIONS ON THE STREAMLAB CORE UNIT. THE OPERATOR TRIPPED ON ONE OF THE PINS THAT PROTRUDE BELOW THE FRAME OF THE INPUT/OUTPUT MODULE AND FELL DOWN. THE OPERATOR BRUISED HER KNEE AND DID NOT REQUIRE MEDICAL TREATMENT. THE OPERATOR REPORTED NO SUSTAINED INJURY DUE TO THIS INCIDENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
271007 STREAMLAB® ANALYTICAL WORKCELL STREAMLAB ANALYTICAL WORKCELL LXG SIEMENS HEALTHCARE DIAGNOSTICS STREAMLAB ANALYTICAL WORKCELL

Patients

Seq Age Sex Outcome Treatment
1