FDA Adverse Event Injury Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3171139 · Received June 14, 2013

Report

Report Number
2032227-2013-02454
Event Type
Injury
Date Received
June 14, 2013
Date of Event
June 10, 2013
Report Date
June 10, 2013
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP ALARMED MOTOR ERROR DURING REWIND DUE TO CORRODED MOTOR HOME SWITCH. UNABLE TO PERFORM THE DISPLACEMENT TEST AND PRIME TEST DUE TO MOTOR ERROR ALARM. THE INSULIN PUMP HAD CRACKED BATTERY TUBE THREADS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP ALARMED MOTOR ERROR DURING THE REWIND. THE INSULIN PUMP WAS NOT EXPOSED TO HIGH MAGNETIC FIELDS. TROUBLESHOOTING WAS PERFORMED, AND THE INSULIN PUMP FAILED THE DISPLACEMENT TEST. NO DAMAGE TO THE DRIVE SUPPORT CAP WAS NOTED. THE INSULIN PUMP PASSED THE SELF TEST. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
271770 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-522LWWB

Patients

Seq Age Sex Outcome Treatment
1