FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3171132 · Received June 14, 2013

Report

Report Number
2032227-2013-02449
Event Type
Malfunction
Date Received
June 14, 2013
Date of Event
May 24, 2013
Report Date
May 24, 2013
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS RECEIVED WITH ALARMS DURING THE BASIC OCCLUSION TEST DUE TO A PROTRUDED/LOOSE DRIVE SUPPORT DISK. THE INSULIN PUMP ALSO HAD A MISSING END CAP STICKER.

Description of Event or Problem · 1

THE CUSTOMER REPORTED AN ALARM DURING THE MANUAL PRIME. THE REPORTED BLOOD GLUCOSE AT THE TIME OF THE CALL WAS 134 MG/DL. ADVISED THE CUSTOMER THAT THE INSULIN PUMP WOULD BE REPLACED. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
270389 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-523LNAL

Patients

Seq Age Sex Outcome Treatment
1 45 YR