FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3171129 · Received June 14, 2013

Report

Report Number
2032227-2013-02456
Event Type
Malfunction
Date Received
June 14, 2013
Date of Event
June 3, 2013
Report Date
June 3, 2013
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP ALARMED MOTOR ERROR DURING THE BASIC OCCLUSION TEST DUE TO A PROTRUDED/LOOSE DRIVE SUPPORT DISK. THE INSULIN PUMP PASSED THE DISPLACEMENT TEST. THE MOTOR ALSO PASSED THE MOTOR TEST.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP ALARMED MOTOR ERROR. THE INSULIN PUMP WAS NOT EXPOSED TO HIGH MAGNETIC FIELDS. TROUBLESHOOTING WAS PERFORMED, AND THE INSULIN PUMP PASSED THE SELF TEST. UNABLE TO CONDUCT THE DISPLACEMENT TEST DUE TO THE MOTOR ERROR ALARMS. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
270295 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-722LWWL