FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3171078 · Received June 14, 2013

Report

Report Number
2032227-2013-02450
Event Type
Malfunction
Date Received
June 14, 2013
Date of Event
May 24, 2013
Report Date
May 24, 2013
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP FAILED THE DISPLACEMENT TEST AND ALARMED MOTOR ERROR DURING THE REWIND DUE TO AN OUT OF PHASE MOTOR ENCODER SIGNAL.

Description of Event or Problem · 1

THE CUSTOMER CALLED TO REPORT THAT SHE WAS GETTING MULTIPLE MOTOR ERROR ALARMS DURING THE PRIME PROCESS. THE CUSTOMER STATED THAT THE INSULIN PUMP WAS NOT EXPOSED TO HIGH MAGNETIC FIELDS. TROUBLESHOOTING WAS PERFORMED, AND CONFIRMED MOTOR ERROR AND OTHER ERROR ALARMS IN THE ALARM HISTORY. ADVISED THE CUSTOMER THAT THE INSULIN PUMP WOULD BE REPLACED. NOTHING FURTHER WAS REPORTED.