PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 3171078
·
Received June 14, 2013
Report
- Report Number
- 2032227-2013-02450
- Event Type
- Malfunction
- Date Received
- June 14, 2013
- Date of Event
- May 24, 2013
- Report Date
- May 24, 2013
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
THE INSULIN PUMP FAILED THE DISPLACEMENT TEST AND ALARMED MOTOR ERROR DURING THE REWIND DUE TO AN OUT OF PHASE MOTOR ENCODER SIGNAL.
Description of Event or Problem · 1
THE CUSTOMER CALLED TO REPORT THAT SHE WAS GETTING MULTIPLE MOTOR ERROR ALARMS DURING THE PRIME PROCESS. THE CUSTOMER STATED THAT THE INSULIN PUMP WAS NOT EXPOSED TO HIGH MAGNETIC FIELDS. TROUBLESHOOTING WAS PERFORMED, AND CONFIRMED MOTOR ERROR AND OTHER ERROR ALARMS IN THE ALARM HISTORY. ADVISED THE CUSTOMER THAT THE INSULIN PUMP WOULD BE REPLACED. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 270825 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-722LNAP |