FDA Adverse Event Malfunction Summary report: N

ACCESS

MDR report key: 3170993 · Received June 14, 2013

Report

Report Number
1416980-2013-15431
Event Type
Malfunction
Date Received
June 14, 2013
Date of Event
May 14, 2013
Report Date
May 22, 2013
Manufacturer
BAXTER HEALTHCARE - SAN JUAN
Product Code
FPA
PMA / PMN Number
K981792
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CI
Reporter Occupation
PHARMACIST

Narratives

Additional Manufacturer Narrative · 1

(B)(4). A BATCH REVIEW WAS CONDUCTED AND REVEALED THAT ALL OF THE ACCEPTANCE CRITERIA WERE MET TO RELEASE THE LOT. THERE WERE NO NON-CONFORMANCES, FAILURES, REWORK, OR DEVIATIONS RELATED TO THE LOT. THERE WERE NO CHANGES TO SPECIFICATIONS, TEST METHODS, PROCESS, EQUIPMENT, OR RAW MATERIALS THAT COULD BE ASSOCIATED WITH THE REPORTED CONDITION. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED. AS THE SAMPLE WAS NOT AVAILABLE, AN EVALUATION COULD NOT BE PERFORMED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT AN INTERLINK PACLITAXEL SET LEAKED FROM THE FILTER AREA DURING INFUSION. THE DEVICE WAS BEING USED WITH AN ONCOLOGICAL DRUG. THERE WAS PATIENT INVOLVEMENT; HOWEVER, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THE REPORTED CONDITION. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
271200 ACCESS SET, ADMINISTRATION, INTRAVASCULAR FPA BAXTER HEALTHCARE - SAN JUAN R12H28117

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN ONCOLOGICAL DRUG