HOMECHOICE PRO
Report
- Report Number
- 1416980-2013-15419
- Event Type
- Malfunction
- Date Received
- June 14, 2013
- Date of Event
- May 17, 2013
- Report Date
- May 24, 2013
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FKX
- PMA / PMN Number
- K102936
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE WAS RETURNED TO BAXTER, AND THE EVALUATION IS COMPLETE. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT OPENED DURING INVESTIGATION OF (B)(4). THE REPORTED DIFFICULTY OF AN IIPV EVENT WAS CONFIRMED THROUGH AN EVENT HISTORY LOG REVIEW. THE CAUSE WAS FALSE EMPTY DETECT AND USE ERROR, INAPPROPRIATE BYPASS OF THE LOW DRAIN VOLUME ALARM, NOT INCLUDING INITIAL DRAIN. HOMECHOICE / HOMECHOICE PRO APD SYSTEMS PATIENT AT-HOME GUIDEGIVES INSTRUCTIONS ON HOW TO BYPASS LDV ALARM. THE WARNING ON PAGE 15-50 STATES "BYPASSING A LOW DRAIN VOLUME ALARM CAN LEAVE FLUID IN THE PERITONEAL CAVITY AND RESULT IN AN INCREASED INTRAPERITONEAL VOLUME (IIPV) SITUATION". IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW-UP WILL BE SENT.
DURING EVALUATION OF A RETURNED HOMECHOICE MACHINE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2013 AT 12:21:38. DURING NIGHT DRAIN CYCLE THREE, THE PATIENT'S ULTRAFILTRATION READING WAS 2230ML, INDICATING THE HOME PATIENT (HP) DRAINED 2230ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 1800ML.THIS INFORMATION MEETS IIPV CRITERIA. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 271466 | HOMECHOICE PRO | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 46 YR |