FDA Adverse Event Malfunction Summary report: N

HOMECHOICE PRO

MDR report key: 3170819 · Received June 14, 2013

Report

Report Number
1416980-2013-15419
Event Type
Malfunction
Date Received
June 14, 2013
Date of Event
May 17, 2013
Report Date
May 24, 2013
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER, AND THE EVALUATION IS COMPLETE. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT OPENED DURING INVESTIGATION OF (B)(4). THE REPORTED DIFFICULTY OF AN IIPV EVENT WAS CONFIRMED THROUGH AN EVENT HISTORY LOG REVIEW. THE CAUSE WAS FALSE EMPTY DETECT AND USE ERROR, INAPPROPRIATE BYPASS OF THE LOW DRAIN VOLUME ALARM, NOT INCLUDING INITIAL DRAIN. HOMECHOICE / HOMECHOICE PRO APD SYSTEMS PATIENT AT-HOME GUIDEGIVES INSTRUCTIONS ON HOW TO BYPASS LDV ALARM. THE WARNING ON PAGE 15-50 STATES "BYPASSING A LOW DRAIN VOLUME ALARM CAN LEAVE FLUID IN THE PERITONEAL CAVITY AND RESULT IN AN INCREASED INTRAPERITONEAL VOLUME (IIPV) SITUATION". IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW-UP WILL BE SENT.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE MACHINE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2013 AT 12:21:38. DURING NIGHT DRAIN CYCLE THREE, THE PATIENT'S ULTRAFILTRATION READING WAS 2230ML, INDICATING THE HOME PATIENT (HP) DRAINED 2230ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 1800ML.THIS INFORMATION MEETS IIPV CRITERIA. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
271466 HOMECHOICE PRO SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1 46 YR