FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3170716 · Received June 14, 2013

Report

Report Number
2531779-2013-08244
Event Type
Malfunction
Date Received
June 14, 2013
Report Date
May 20, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED TO ANIMAS. EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 07/15/2013 WITH THE FOLLOWING FINDINGS: A VISUAL INSPECTION OF THE PUMP CONFIRMED EVIDENCE OF MOISTURE IN THE DISPLAY LENS. DURING TESTING, THE PUMP WAS UNABLE TO POWER ON. A CRACK WAS FOUND ON THE BATTERY COMPARTMENT AND THERE WAS VISIBLE CORROSION INSIDE. THE PUMP FAILED AN IN-HOUSE LEAK TEST AT THE BATTERY COMPARTMENT. THE PUMP COVER WAS REMOVED AND INTERNAL MOISTURE DAMAGE WAS OBSERVED. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (MOISTURE INGRESS) ISSUE. THE REPORTER STATED THAT THE PUMP WOULD NOT POWER BACK ON AFTER A BATTERY CHANGE AFTER BEING EXPOSED TO WATER. THE REPORTER NOTED MOISTURE BEHIND THE DISPLAY SCREEN AND DENIED ANY DAMAGE TO THE BATTERY CAP OR BATTEY COMPARTMENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
271794 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 32 YR