FDA Adverse Event Malfunction Summary report: N

OLYMPUS RESECTION SHEATH

MDR report key: 3170658 · Received March 22, 2013

Report

Report Number
9610773-2013-00036
Event Type
Malfunction
Date Received
March 22, 2013
Date of Event
May 2, 2013
Report Date
May 6, 2013
Manufacturer
OLYMPUS WINTER & IBE GMBH
Product Code
FED
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

OLYMPUS FOLLOWED-UP WITH THE USER FACILITY TO OBTAIN ADDITIONAL INFORMATION REGARDING THIS REPORT. THE USER FACILITY REPORTED THAT THE TIP BROKE OFF INSIDE THE PATIENT'S BLADDER DURING A THERAPEUTIC TRANSURETHRAL RESECTION OF THE PROSTATE (TURP) PROCEDURE WHERE THE TIP WAS RETRIEVED WITH AN UNSPECIFIED GRASPER. THE INTENDED PROCEDURE REPORTEDLY WAS ABORTED AND HAD NOT YET BEEN RESCHEDULED. THE DEVICE REFERENCED IN THIS REPORT WAS RETURNED TO OLYMPUS FOR EVALUATION. THE EVALUATION CONFIRMED THE USER'S REPORT. THE INSULATION BEAK WAS CRACKED AND BROKE OFF FROM THE METAL SHEATH AT THE DISTAL END. THERE WERE NO SIGNS OF INDENTATION OR DAMAGES NOTED ON THE METAL SHEATH. THE INSULATION BEAK WAS AN OLDER VERSION. THE REFERENCED DEVICE HAD BEEN FORWARDED TO THE OEM FOR FURTHER EVALUATION.

Description of Event or Problem · 1

OLYMPUS WAS INFORMED THAT THE REFERENCED DEVICE FAILED AFTER SEVERAL USES. THE TIP FELL OFF DURING AN UNSPECIFIED PROCEDURE. THERE WAS NO PATIENT INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
121083 OLYMPUS RESECTION SHEATH RESECTION SHEATH FED OLYMPUS WINTER & IBE GMBH A22041A 125W

Patients

Seq Age Sex Outcome Treatment
1 57 YR OLYMPUS UES-40 GENERATOR WITH UNK SERIAL #