OMNIPOD INSULIN PUMP
Report
- Report Number
- 3004464228-2013-00499
- Event Type
- Malfunction
- Date Received
- May 28, 2013
- Date of Event
- April 28, 2013
- Report Date
- April 28, 2013
- Manufacturer
- INSULET CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042792
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE DEVICE WAS NOT RETURNED FOR EVALUATION. WE ARE UNABLE TO CONFIRM THE REPORTED FAILURE OF THE CANNULA TO DEPLOY OR TO DETERMINE ITS ROOT CAUSE. THE OMNIPOD USER GUIDE WARNS "CHECK THE INFUSION SITE AFTER INSERTION TO ENSURE THAT THE CANNULA WAS PROPERLY INSERTED. IT IS ALSO A GOOD IDEA TO CHECK YOUR BLOOD GLUCOSE ABOUT TWO HOURS AFTER EACH POD CHANGE AND TO CHECK THE INFUSION SITE PERIODICALLY. IF THE CANNULA IS NOT PROPERLY INSERTED, HYPERGLYCEMIA MAY RESULT," AND "TEST RESULTS GREATER THAN 250 MG/DL MEAN HIGH BLOOD GLUCOSE (HYPERGLYCEMIA). IF YOU GET RESULTS ABOVE 250 MG/DL, BUT DO NOT HAVE SYMPTOMS OF HYPERGLYCEMIA, REPEAT THE TEST. IF YOU HAVE SYMPTOMS OR CONTINUE TO GET RESULTS THAT FALL ABOVE 250 MG/DL, FOLLOW THE TREATMENT ADVICE OF YOUR HEALTHCARE PROVIDER." QUALIFICATION RECORDS REVIEWED AND THE PRODUCT LOT MET ALL ACCEPTANCE CRITERIA.
THE CUSTOMER REPORTED THAT HIS BLOOD GLUCOSE REACHED 270 MG/DL. HE GAVE A MANUAL INJECTION FOR CORRECTION AND REMOVED THE DEVICE. HE REALIZED THAT THE CANNULA WAS STILL INSIDE THE POD AND HAD NEVER INSERTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 233152 | OMNIPOD INSULIN PUMP | PUMP, INFUSION, INSULIN | LZG | INSULET CORPORATION | 11200 | L31102 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 62 YR |