FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 4000 SYSTEM
MDR report key: 3170362
·
Received May 24, 2013
Report
- Report Number
- 3008642652-2013-01407
- Event Type
- Malfunction
- Date Received
- May 24, 2013
- Date of Event
- May 9, 2013
- Report Date
- May 21, 2013
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (CHECK ELECTRODE BELT MESSAGES) WAS CONFIRMED. UPON INVESTIGATION THE ELECTRODE BELT FAILED A PULSE LEAD HI-POT TEST AND A THERAPY ELECTRODE RECOGNITION TEST. THE CAUSE FOR THE FAILURES WAS ISOLATED TO AN OPEN PULSE WIRE IN THE TRUNK CABLE CONNECTOR. THE ROOT CAUSE FOR THE OPEN WIRE COULD NOT BE POSITIVELY IDENTIFIED BUT WAS LIKELY EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE OPEN PULSE WIRE. THE PATIENT RECEIVED A REPLACEMENT ELECTRODE BELT.
Description of Event or Problem · 1
A (B)(6) MALE PATIENT CALLED ZOLL CUSTOMER SUPPORT TO REPORT THAT HE WAS RECEIVING CHECK ELECTRODE BELT MESSAGES. THE PATIENT WAS ISSUED A REPLACEMENT ELECTRODE BELT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 231055 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 71 YR |