FDA Adverse Event Injury Summary report: N

STANDARD CRANIOTOME, BLACK MAX

MDR report key: 3170280 · Received June 14, 2013

Report

Report Number
1045834-2013-02463
Event Type
Injury
Date Received
June 14, 2013
Date of Event
April 29, 2013
Report Date
May 14, 2013
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
GFF
PMA / PMN Number
K974025
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

AS OF THIS DATE, THE DEVICE HAS NOT BEEN RETURNED FOR EVALUATION; THEREFORE, THE REPORTED CONDITION CANNOT BE CONFIRMED AND/OR DUPLICATED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SENT ACCORDINGLY. (B)(4).

Additional Manufacturer Narrative · 1

THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. RELIABILITY ENGINEERING EVALUATED THE DEVICE. IT WAS OBSERVED THAT THE BACK POST AND PROTECTIVE FOOT WERE DRILLED INTO AND PROTECTIVE FOOT WAS BROKEN OFF AT THE DRILLED LOCATION. THEREFORE, THE REPORTED CONDITION WAS CONFIRMED. EVIDENCE SUGGESTS THIS WAS DUE TO MISUSE AT THE CUSTOMER SITE. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SENT ACCORDINGLY. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A CRANIOTOMY PROCEDURE, THE TIP OF THE ATTACHMENT DEVICE "BROKE OFF AND PENETRATED THE DURA MATER AT THE TRANSVERSE SINUS OF THE PATIENT'S CRANIUM." THE PLANNED SURGICAL PROCEDURE WAS DELAYED; HOWEVER, THE LENGTH OF THE DELAY WAS UNKNOWN TO THE REPORTER. A TRANSIENT INJURY TO THE PATIENT WAS ALLEGED, WHICH NECESSITATED ADDITIONAL MEDICAL TREATMENT TO REMEDIATE THE INJURY. THE TYPE OF ADDITIONAL TREATMENT REQUIRED WAS UNKNOWN TO THE REPORTER. THE POST-OPERATIVE STATUS OF THE PATIENT IS UNKNOWN TO THE REPORTER AT THIS TIME. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
272193 STANDARD CRANIOTOME, BLACK MAX BUR, SURGICAL, GENERAL & PLASTIC SURGERY - ATTACHMENT GFF DEPUY SYNTHES POWER TOOLS

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R