MEGASUTURECUT NEEDLE DRIVER INSTRUMENT
Report
- Report Number
- 2955842-2013-02169
- Event Type
- Malfunction
- Date Received
- June 14, 2013
- Date of Event
- May 22, 2013
- Report Date
- June 19, 2013
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THAT ONE GRIP CLOSE CABLE WAS FOUND TO BE BROKEN AT THE DISTAL IDLERS. THE IDLER PULLEY WAS ABLE TO SPIN FREELY AND IT DID NOT EXHIBIT ANY DAMAGE. THE CABLE SEGMENT STUCK OUT AT THE WRIST. OTHER CABLES AT WRIST WERE NOT DAMAGED. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A REPORTABLE EVENT; HOWEVER, THE BROKEN CABLE IF TO REOCCUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.
THE INSTRUMENT HAS NOT BEEN RETURNED FOR EVALUATION; THEREFORE, THE ROOT CAUSE OF THE CUSTOMER REPORTED FAILURE MODE CANNOT BE DETERMINED. A FOLLOW-UP MDR WILL BE SUBMITTED IF THE INSTRUMENT IS RETURNED (POST ENGINEERING EVALUATION) OR IF ADDITIONAL INFORMATION IS RECEIVED. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A REPORTABLE EVENT; HOWEVER, THE BROKEN WIRES IF TO REOCCUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.
IT WAS REPORTED THAT DURING A DA VINCI S HYSTERECTOMY PROCEDURE, THE CUSTOMER NOTED BROKEN WIRES ON THE ON THE MEGA SUTURE CUT NEEDLE DRIVER. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 270626 | MEGASUTURECUT NEEDLE DRIVER INSTRUMENT | ENDOSCOPIC INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420309-01 | M10120704 106 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | DA VINCI S SYSTEM INSTRUMENTS AND ACCESSORIES |