FDA Adverse Event Injury Summary report: N

ENDO GIA ROTICULATOR 45-2.0 SULU

MDR report key: 3170119 · Received June 6, 2013

Report

Report Number
1219930-2013-00429
Event Type
Injury
Date Received
June 6, 2013
Date of Event
May 14, 2013
Report Date
May 16, 2013
Manufacturer
CO IDIEN, FORMERLY US SUR
Product Code
GAG
PMA / PMN Number
K061095
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

PROCEDURE: VATS/RIGHT LOWER LOBECTOMY. ACCORDING TO THE REPORTER: ROTICULATOR 45-3.5 SULU WAS USED TO FORM INTERLOBAR AND ROTICULATOR 45-2.5 SULU WAS USED TO CUT THE PULMONARY ARTERY. AFTER 3RD FIRING ON THE INFERIOR PULMONARY VEIN WITH ROTICULATOR 40-2.0 SULU, BLOOD SPURTED FROM AROUND THE CENTER OF THE STUMP. THE SMALL INCISION WAS EXTENDED BY ABOUT 3 CM, AND THE BLEEDING SITE WAS GRASPED WITH THE FORCEPS AND WAS CAREFULLY SUTURED. BLEEDING STOPPED BEFORE IT BECAME SERIOUS. AMOUNT OF BLEEDING: ABOUT 300 CC. THERE WAS TISSUE DAMAGE. NO UNANTICIPATED TISSUE LOSS, NOTHING FELL INTO CAVITY, THE PT IS RECOVERING. OPERATING TIME WAS EXTENDED BY MORE THAN 2 HOURS. NO INFORMATION ABOUT THE USE OF REINFORCEMENT MATERIAL WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
251946 ENDO GIA ROTICULATOR 45-2.0 SULU DISPSOABLE SURGICAL STAPLER GAG CO IDIEN, FORMERLY US SUR N2K0401LX

Patients

Seq Age Sex Outcome Treatment
1 Other GIA UNIVERSAL, 030403, LOT NUMBER: N2J0409X| EXP DATE: 09/30/2017, MANUF DATE: 09/01/2012