LOCKSCR Ø2.4 SELF-TAP L16 TAN
Report
- Report Number
- 2520274-2013-02913
- Event Type
- Malfunction
- Date Received
- June 14, 2013
- Date of Event
- May 9, 2013
- Report Date
- May 15, 2013
- Manufacturer
- SYNTHES USA
- Product Code
- HRS
- PMA / PMN Number
- K102694
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SZ
- Reporter Occupation
- OTHER
Narratives
DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. SUBJECT DEVICE HAS BEEN RECEIVED AND IS CURRENTLY IN THE EVALUATION PROCESS. INVESTIGATION IS ON GOING; NO CONCLUSION COULD BE DRAWN. WITHOUT A LOT NUMBER THE DEVICE HISTORY RECORDS REVIEW COULD NOT BE COMPLETED.
AN ADDITIONAL EVALUATION WAS PERFORMED BY SYNTHES (B)(4) AND THE REPORT INDICATES: VISUAL INSPECTION SHOWS THE SURFACE HAVE DISCOLOURED AREAS. THIS IS INDICATIVE OF DEVICE WAS SUBJECT TO MULTIPLE WASHING PROCESSES/CYCLES WHICH LED TO AN ACCELERATED DISCOLORATION OF THE DEVICE. THIS IS ALSO DUE TO DEVICE IS MADE OF ANODISING TITANIUM. THAT MEANS THIS IS AN OXIDATION WITHIN M AREA. ACCORDING OF THE THICKNESS OF THIS OXIDATION COATING THE VISIBLE COLOUR IS CHANGING. IF THIS COATING IS CHANGING BECAUSE OF SOME CIRCUMSTANCES LIKE SPLASHES WITH A LIQUID, THE COLOUR WILL ALSO CHANGE. ALTHOUGH DISCOLORED, THE FUNCTIONALITY OF THE DEVICE(S) REMAINS INTACT.
DEVICE REPORT FROM SYNTHES (B)(6) REPORTS AN EVENT IN (B)(6) AS FOLLOWS: PATIENT WAS IMPLANTED WITH VA-LCP 2 COLUMN DISTAL RADIUS PLATE AND SCREW CONSTRUCT ON (B)(6) 2011. REPORTEDLY THE HEALING PROCESS WAS COMPLETED. PATIENT WAS RETURNED TO THE OR ON (B)(6) 2013 FOR REMOVAL OF HARDWARE. DURING THE IMPLANT REMOVAL, IT WAS NOTED THAT THE COLOR OF THE VA LOCKING SCREWS CHANGED AND THE SCREWS APPEARED TO HAVE CORROSION. THIS IS 4 OF 9 REPORTS FOR THE SAME EVENT, COMPLAINT (B)(4).
THIS IS REPORT 4 OF 9 FOR COMPLAINT (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 272074 | LOCKSCR Ø2.4 SELF-TAP L16 TAN | HRS | SYNTHES USA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 57 YR |