DIMENSION EXL WITH LM
Report
- Report Number
- 1226181-2013-00262
- Event Type
- Malfunction
- Date Received
- June 13, 2013
- Date of Event
- May 22, 2013
- Report Date
- May 22, 2013
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS
- Product Code
- JJE
- PMA / PMN Number
- K130276
- Removal / Correction Number
- 1226181-07/16/2014-003-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- OTHER
Narratives
A SIEMENS FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE. THE FSE ANALYZED THE INSTRUMENT AND DETERMINED THAT THE CAUSE OF THE INSTRUMENT COVER FALLING DOWN WAS DUE TO A FAULTY COVER HINGE. THE FSE ADJUSTED THE COVER HINGE TENSION AND THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. FURTHER EVALUATION OF THE DEVICE IS NOT REQUIRED.
THE INITIAL MDR 1226181-2013-00262 WAS FILED ON JUNE 13, 2013. ADDITIONAL INFORMATION (07/03/2014): SIEMENS HEALTHCARE DIAGNOSTICS HAS INVESTIGATED THE INSTANCES OF DIMENSION EXL INSTRUMENT REAGENT LIDS FALLING DURING MAINTENANCE PROCEDURES. IT HAS BEEN DETERMINED THAT THE HINGE MAY LOSE ITS EFFECTIVENESS AND SLOWLY SHIFT DOWNWARD DURING MAINTENANCE PROCEDURES. CUSTOMERS IN THE UNITED STATES WERE SENT URGENT MEDICAL DEVICE CORRECTION (UMDC) 14-46 ENTITLED "DIMENSION EXL REAGENT LID HINGE ISSUE" AND CUSTOMERS OUTSIDE OF THE UNITED STATES WERE SENT URGENT FIELD SAFETY NOTICE (UFSN) 14-46 ENTITLED "DIMENSION EXL REAGENT LID HINGE ISSUE". THE UMDC/UFSN INFORM CUSTOMERS THAT THEIR CSE WILL INSPECT AND ADJUST THE TENSION ON THE REAGENT LID HINGES DURING EVERY VISIT. IF THE HINGES CAN NO LONGER BE PROPERLY ADJUSTED, THE HINGES WILL BE REPLACED. THE UMDC/UFSN ALSO INSTRUCTS CUSTOMERS TO CONTACT THE SIEMENS CUSTOMER CARE CENTER OR THEIR LOCAL SIEMENS TECHNICAL SUPPORT REPRESENTATIVE IF THEY OBSERVE THAT THE REAGENT LID IS SHIFTING DOWNWARD WHILE IN THE OPEN POSITION. THE INITIAL MDR LISTED THE 510(K) NUMBER AS K073604. THE CORRECT 510(K) NUMBER FOR DIMENSION EXL W/LM INSTRUMENTS IS K130276.
THE OPERATOR OF AN DIMENSION EXL WITH LM ANALYZER WAS PEFORMING ROUTINE WORK WHEN THE COVER OF THE ANALYZER FELL DOWN AND CAME IN CONTACT WITH HER HEAD. THE CUSTOMER SOUGHT MEDICAL ATTENTION AND WAS EVALUATED AT THE EMERGENCY ROOM. THE PHYSICIAN ORDERED A CAT SCAN AND DETERMINED THAT THE OPERATOR ENCOUNTERED NO INJURIES DUE TO THE INSTRUMENT COVER CONTACTING HER HEAD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 270211 | DIMENSION EXL WITH LM | CLINICAL CHEMISTRY SYSTEM | JJE | SIEMENS HEALTHCARE DIAGNOSTICS | DIMN EXL WITH LM |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |