FDA Adverse Event
Malfunction
Summary report: N
OUTLOOK
MDR report key: 3167714
·
Received June 11, 2013
Report
- Report Number
- 1641965-2013-00032
- Event Type
- Malfunction
- Date Received
- June 11, 2013
- Date of Event
- March 19, 2013
- Report Date
- May 14, 2013
- Manufacturer
- B. BRAUN MEDICAL, INC.
- Product Code
- FRN
- PMA / PMN Number
- K011975
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
THE REPORTED COMPLAINT WAS CONFIRMED. SOLENOID HAD OVERHEATED. REPLACED SOLENOID #113007, LEFT STANDOFF #401542, RIGHT STANDOFF #401543. PERFORMED PREVENTATIVE MAINTENANCE SERVICE.
Description of Event or Problem · 1
PAIN MEDICATION NEEDED POST SURGERY WAS NOT BEING DELIVERED TO A PT DUE TO THE OUTLOOK 100 PUMP FREQUENTLY DISPLAYING AN EMPTY BAG ALARM. THE PT WAS IN PAIN DUE TO THE TREATMENT BEING DELAYED. THE NURSE HAD TO FREQUENTLY RESTART THE PUMP
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 263901 | OUTLOOK | INFUSION PUMP | FRN | B. BRAUN MEDICAL, INC. | OUTLOOK 100 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |