PINNACLE MTL INS NEUT36IDX58OD
Report
- Report Number
- 1818910-2013-18809
- Event Type
- Injury
- Date Received
- June 13, 2013
- Date of Event
- June 7, 2013
- Report Date
- December 16, 2013
- Manufacturer
- DEPUY ORTHOPAEDICS INC US
- Product Code
- KWA
- PMA / PMN Number
- PK003523
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE DEVICES ASSOCIATED WITH THIS REPORT WERE NOT RETURNED. A REVIEW OF THE DEVICE HISTORY RECORDS FOR LOT CODES 1951481 AND 1846927 DID NOT REVEAL ANY RELATED MANUFACTURING ANOMALIES. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY PRODUCT CONTRIBUTION TO THE REPORTED EVENT WITH THE INFORMATION PROVIDED. BASED ON THE INABILITY TO DETERMINE A ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.
THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.
UPDATE REC'D (B)(4) 2013 - PFS AND MEDICAL RECORDS RECEIVED. PART/LOT WAS PROVIDED. REVISION OPERATIVE NOTES INDICATE CORROSION. THE STEM HAS NOW BEEN REPORTED. THE INVESTIGATION IS ONGOING. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.
**UPDATE: (B)(4) 2013 - LITIGATION PAPERS RECEIVED. LITIGATION ALLEGES THAT THE PATIENT SUFFERS FROM DISCOMFORT, INFLAMMATION, AND LARGE AMOUNTS OF TOXIC COBALT CHROMIUM METAL ION PARTICLES TO BE RELEASED INTO THE BLOOD, TISSUE, AND BONE. THE INVESTIGATION IS ONGOING. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.
THE DEVICES ASSOCIATED WITH THIS REPORT WERE NOT RETURNED. A COMPLAINT DATABASE SEARCH FOR THE ADDED STEM DID NOT FIND ANY OTHER COMPLAINTS AGAINST THE PROVIDED LOT CODE. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY PRODUCT CONTRIBUTION TO THE REPORTED EVENT WITH THE INFORMATION PROVIDED. BASED ON THE INABILITY TO DETERMINE A ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.
(B)(4). NO DEVICE ASSOCIATED WITH THIS REPORT WAS RECEIVED FOR EXAMINATION. A WORLDWIDE LOT SPECIFIC COMPLAINT DATABASE SEARCH, OR DEVICE HISTORY RECORD (DHR) REVIEW WAS NOT POSSIBLE BECAUSE THE REQUIRED LOT CODE(S) WAS NOT PROVIDED. BASED ON PREVIOUS INVESTIGATIONS THIS COMPLICATION OF JOINT REPLACEMENT IS UNLIKELY TO HAVE BEEN THE RESULT OF A DEVICE FAILING TO MEET REQUIRED SPECIFICATIONS. THE INFORMATION RECEIVED WILL BE RETAINED FOR POTENTIAL SERIES INVESTIGATIONS IF TRIGGERED BY TREND ANALYSIS, POST MARKET SURVEILLANCE, OR OTHER EVENTS WITHIN THE QUALITY SYSTEM. CORRECTIVE ACTION WAS NOT INDICATED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH , A FOLLOW-UP MEDWATCH, WILL BE FILED AS APPROPRIATE.
REVISION DUE TO PAIN AND POSSIBLE METAL ON METAL REACTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 269183 | PINNACLE MTL INS NEUT36IDX58OD | ACETABULAR CUP | KWA | DEPUY ORTHOPAEDICS INC US | 1846927 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 54 YR | Required Intervention |