ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-08111
- Event Type
- Malfunction
- Date Received
- June 11, 2013
- Report Date
- May 16, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
FOLLOW-UP # 1 DATE OF SUBMISSION 08/06/2013 - DEVICE EVALUATION:THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 07/15/2013 WITH THE FOLLOWING FINDINGS: THE KEYPAD BUTTON SYMBOLS WERE FOUND TO BE WORN OFF; HOWEVER THE KEYPAD WAS FULLY INTACT. DURING TESTING, THE UP ARROW, DOWN ARROW AND CONTRAST KEYPAD BUTTONS WERE FOUND TO BE INTERMITTENTLY UNRESPONSIVE AND REQUIRED MULTIPLE BUTTON PRESSES TO ELICIT A RESPONSE; THE OK KEYPAD BUTTON RESPONDED APPROPRIATELY. THERE WAS EVIDENCE OF CONTAMINATION FOUND UNDER ALL KEY CONTACTS.
THE PUMP HAS BEEN RETURNED TO ANIMAS FOR EVALUATION; HOWEVER, PRODUCT INVESTIGATION HAS NOT BEEN COMPLETED. ONCE EVALUATION HAS BEEN COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (TACTILE CHANGES/UNRESPONSIVE) ISSUE. REPORTEDLY, THE UP AND DOWN BUTTONS ARE NOT RESPONDING CONSISTENTLY TO USER INPUT FOR THE LAST 4 WEEKS.. THE SYMBOLS ON THE KEYPAD IS WORN. THERE IS NO EVIDENCE OF PRODUCT MISUSE, MOISTURE, OR CORROSION ASSOCIATED WITH THE REPORTED ISSUE. THE PUMP IS BEING RETURNED FOR INVESTIGATION. THERE IS NO INDICATION THAT THE PRODUCT ISSUE CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ISSUE REMAINED UNRESOLVED AT THE TIME OF TROUBLESHOOTING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 263760 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 13 YR |