FDA Adverse Event Malfunction Summary report: N

LARGE NEEDLE DRIVER INSTRUMENT

MDR report key: 3163542 · Received June 11, 2013

Report

Report Number
2955842-2013-02107
Event Type
Malfunction
Date Received
June 11, 2013
Date of Event
December 14, 2012
Report Date
May 17, 2013
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THAT THE PITCH UP CABLE WAS FOUND TO BE FRAYED AT THE DISTAL CLEVIS HUB. THE FRAYED STRANDS STUCK OUT AT THE WRIST. OTHER CABLES AT WRIST WERE NOT DAMAGED. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A REPORTABLE EVENT; HOWEVER, THE FRAYED CABLE IF TO REOCCUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI SI PROSTATECTOMY PROCEDURE, THE CUSTOMER NOTED A LOOSE WIRE ON THE LARGE NEEDLE DRIVER INSTRUMENT. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
263944 LARGE NEEDLE DRIVER INSTRUMENT ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420006-06 M10121005 910

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI SI SYSTEM INSTRUMENTS AND ACCESSORIES