FDA Adverse Event Malfunction Summary report: N

ADULT DUAL HEATED EVAQUA BREATHING CIRCUIT

MDR report key: 3163346 · Received June 11, 2013

Report

Report Number
9611451-2013-00458
Event Type
Malfunction
Date Received
June 11, 2013
Date of Event
May 7, 2013
Report Date
May 13, 2013
Manufacturer
FISHER & PAYKEL HEALTHCARE LTD
Product Code
BZE
PMA / PMN Number
K983112
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). METHOD: THE COMPLAINT BREATHING CIRCUIT WAS RETURNED TO FISHER & PAYKEL HEALTHCARE (B)(4) AND VISUALLY INSPECTED. RESULTS: THE VISUAL INSPECTION REVEALED A HOLE ON THE EXPIRATORY EVAQUA LIMB, APPROXIMATELY 29CM FROM THE ELBOW CONNECTOR. A LOT CHECK COULD NOT BE PERFORMED AS NO LOT NUMBER WAS PROVIDED. CONCLUSION: BASED ON THE INSPECTION OF THE DAMAGE, IT IS LIKELY THAT THE EVAQUA EXPIRATORY LIMB WAS PUNCTURED OR SCRATCHED BY A BLUNT OBJECT. ALL RT340 BREATHING CIRCUITS ARE PRESSURE TESTED FOR LEAKS PRIOR TO BEING RELEASED FOR DISTRIBUTION. IN ADITION, TUBING WEIGHT AND BOND STRENGTH TESTS ARE PERFORMED EVERY 15 MINUTES. ANY BREATHING CIRCUIT WHICH FAILS ANY OF THESE TESTS IS DISCARDED. THIS SUGGESTS THAT THE BREATHING CIRCUIT WAS DAMAGED AFTER IT WAS RELEASED FOR DISTRIBUTION. THE KEY DIFFERENCE BETWEEN THE EVAQUA BREATHING CIRCUITS AND CONVENTIONAL BREATHING CIRCUITS IS THAT THE EXPIRATORY TUBE OF EVAQUA CIRCUITS SUCH AS THE RT340 IS COMPOSED OF A THIN, SEMI-PERMEABLE FILM SPECIALLY DESIGNED TO ALLOW WATER VAPOUR FROM EXPIRED VENTILATORY GAS TO PASS THROUGH. THE EVAQUA EXPIRATORY TUBE HAS A PROTECTIVE MESH WHICH PREVENTS DAMAGE TO THE WALLS OF THE TUBE, HOWEVER THE EVAQUA TUBING REMAINS MORE SUSCEPTIBLE TO DAMAGE THAN CONVENTIONAL CIRCUITS WHEN EXPOSED TO ROUGH HANDLING OR DAMAGE CAUSED BY SHARP OBJECTS AND NON-FPH CIRCUIT HANGERS. THE USER INSTRUCTIONS SUPPLIED WITH THE RT340 BREATHING CIRCUIT STATE: PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT; SET APPROPRIATE VENTILATOR ALARMS; FIT ONLY THE SUPPLIED FISHER & PAYKEL HEALTHCARE CIRCUIT HANGER WITH CARE TO AVOID CIRCUIT DAMAGE.

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED VIA OUR DISTRIBUTOR THAT THERE WAS A LEAK IN THE EXPIRATORY LIMB OF AN RT340 ADULT DUAL HEATED EVAQUA BREATHING CIRCUIT. NO PATIENT CONSEQUENCE WAS REPORTED.