FDA Adverse Event Malfunction Summary report: N

LIGAMAX-5MM ENDO CLIP APPLIER

MDR report key: 3163209 · Received June 11, 2013

Report

Report Number
3005075853-2013-02934
Event Type
Malfunction
Date Received
June 11, 2013
Report Date
May 20, 2013
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
FZP
PMA / PMN Number
K050344
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TW
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RETURNED IN GOOD VISUAL CONDITION AND WITH A PHOTO WITH ONE CLIP WITH GAP AND ONE MALFORMED CLIP. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON TESTING, THE DEVICE WAS CYCLED, FED, AND FORMED THE REMAINING CLIPS AS INTENDED. IN ORDER TO CONFIRM THE CLIPS WERE WITHIN MANUFACTURING SPECIFICATIONS, THE CLIPS WERE EVALUATED USING A TOOL THAT IS DESIGNED TO DETERMINE PROPER FORMATION. IN ADDITION, THE DEVICE LOCKED OUT AS INTENDED. PLEASE NOTE THAT PRIOR TO LOADING A CLIP IN THE JAWS, ENSURE THAT THE DEMARCATION BETWEEN THE JAWS AND THE DEVICE SHAFT IS PAST THE END OF THE TROCAR CANNULA. ADDITIONALLY, EXCESSIVELY APPLYING A SIDE LOAD TO THE JAWS, CAUSING THEM TO PARTIALLY COLLAPSE COULD RESULT IN A CLIP MALFORMATION. THE DEVICE JAWS SHOULD BE FULLY OPEN AND PARALLEL UPON INITIATING THE FIRING OF THE DEVICE. THE BATCH HISTORY RECORDS WERE REVIEWED WITH NO ANOMALIES NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE CLIP CANNOT BE CORRECTLY FORMED INTO THE SHAPE IT¿S SUPPOSED TO BE (BENDED WHILE FIRING). ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
264154 LIGAMAX-5MM ENDO CLIP APPLIER CLIP, IMPLANTABLE FZP ETHICON ENDO-SURGERY, LLC. NA J4AL35

Patients

Seq Age Sex Outcome Treatment
1