OT VERIO METER
Report
- Report Number
- 3008382007-2013-16676
- Event Type
- Malfunction
- Date Received
- June 11, 2013
- Report Date
- June 4, 2013
- Manufacturer
- LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
- Product Code
- NBW
- PMA / PMN Number
- K093745
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- OTHER
Narratives
FOLLOW-UP # 1 ¿ (09/22/2013). THE PATIENT¿S METER HAS BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS ON 9/10/2013 AND 9/14/2013, RESPECTIVELY, WITH THE FOLLOWING FINDINGS:THE METER PASSED ALL TESTING WITH NO FAULTS FOUND. THE REPORTED ISSUE COULD NOT BE REPRODUCED. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.
ON (B)(6) 2013 THE LAY USER/ PATIENT CONTACTED LIFESCAN (LFS) (B)(4) ALLEGING THE ONETOUCH VERIO METER READ INACCURATELY COMPARED TO ANOTHER DEVICE. THE PATIENT REPORTED BLOOD GLUCOSE RESULTS OF ¿9.7MMOL/L¿ WITH THE SUBJECT METER AND ¿6.9 MMOL/L¿ ON ANOTHER METER, PERFORMED WITHIN 30 MINUTES OF EACH OTHER. THE PATIENT DID NOT EXPERIENCE ANY SYMPTOMS OR SEEK ANY MEDICAL ATTENTION BECAUSE OF THE REPORTED ISSUE. AS THE RESULTS FELL OUTSIDE EXPECTED VALUES FOR METER TO METER ACCURACY TESTING, THIS COMPLAINT IS BEING REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 263783 | OT VERIO METER | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL | 3380342 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR |