FDA Adverse Event Malfunction Summary report: N

AFFINITY OXY NT 511

MDR report key: 3163079 · Received June 11, 2013

Report

Report Number
2184009-2013-00031
Event Type
Malfunction
Date Received
June 11, 2013
Date of Event
April 4, 2013
Report Date
April 8, 2013
Manufacturer
MEDTRONIC PERFUSION SYSTEMS
Product Code
DTZ
PMA / PMN Number
K973760
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT AT MEDTRONIC'S QUALITY LABORATORY, VISUAL INSPECTION DID NOT SHOW ANY OUTWARD SIGNS OF CRACKS OR DAMAGE THROUGHOUT OXY OR PORTS. PERFORMANCE ANALYSIS: BLOOD SIDE PRESSURE INTEGRITY TESTING WAS PERFORMED AT 3 LITERS/MINUTE WITH 23 PSIG OF BACK PRESSURE FOR 10 MINUTES. DURING THE TEST, THERE WAS A LEAK OBSERVED FROM THE HEX (HEAT EXCHANGER) SIDE SEAM. THE DEVICE HISTORY RECORD WAS REVIEWED AND SHOWED THAT THIS PRODUCT MET ALL MANUFACTURING SPECIFICATIONS FOR PRODUCT RELEASED FOR DISTRIBUTION. NO ISSUES WERE IDENTIFIED THAT WOULD HAVE IMPACTED THIS EVENT. CONCLUSION: BASED ON ANALYSIS AND INVESTIGATION, THE LEAK WAS CONFIRMED TO ORIGINATE FROM THE HEX SIDE SEAM. A PARTIAL VOID MAY HAVE FORMED DURING ADHESIVE DISPENSING INTO THE ADHESIVE GROOVE OF THE HEAT EXCHANGER, ALLOWING FOR ACCEPTABLE PRESSURE TEST RESULTS PRIOR TO DISTRIBUTION. IT IS POSSIBLE A PHYSICAL SHOCK ENCOUNTERED DURING SHIPPING OR HANDLING OF THE PRODUCT MAY HAVE COMPROMISED THAT BOND. HISTORICALLY, IT HAS BEEN OBSERVED THAT OVERLY AGGRESSIVE SHIPPING AND HANDLING CAN CAUSE THE PARTIAL BOND TO BECOME COMPROMISED, RESULTING IN A LEAK AND CONTRIBUTING TO THE EVENT. MEDTRONIC IS MONITORING FOR SIMILAR COMPLAINTS. (B)(4).

Description of Event or Problem · 1

MEDTRONIC RECEIVED INFORMATION THAT DURING PRIMING, A LEAK WAS FOUND NEAR THE HEAT EXCHANGER. THE EXACT LOCATION OF THE LEAK WAS NOT SPECIFIED. THE OXYGENATOR WAS CHANGED OUT WITH NO ADVERSE PATIENT EFFECTS REPORTED. THE PRODUCT WAS RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
263779 AFFINITY OXY NT 511 OXYGENATOR, CARDIOPULMONARY BYPASS DTZ MEDTRONIC PERFUSION SYSTEMS 95212

Patients

Seq Age Sex Outcome Treatment
1 Other