FDA Adverse Event Malfunction Summary report: N

OT VERIO METER

MDR report key: 3163048 · Received June 11, 2013

Report

Report Number
3008382007-2013-16674
Event Type
Malfunction
Date Received
June 11, 2013
Report Date
May 29, 2013
Manufacturer
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
Product Code
NBW
PMA / PMN Number
K093745
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 ¿ (09/02/2013). THE PATIENT¿S METER AND TEST STRIPS HAVE BEEN RETURNED ON 8/13/2013 AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS ON 8/28/2013 AND 8/29/2013, RESPECTIVELY, WITH THE FOLLOWING FINDINGS:THE METER PASSED ALL TESTING WITH NO FAULTS FOUND. THE REPORTED ISSUE COULD NOT BE REPRODUCED. THE TEST STRIPS WERE ALSO TESTED AND IT WAS FOUND TO HAVE RESULTS ABOVE RANGE WHEN TESTED WITH CONTROL SOLUTION. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.

Description of Event or Problem · 1

ON (B)(6) 2013, THE LAY USER/ PATIENT CONTACTED LIFESCAN (LFS) (B)(6) ALLEGING THE ONETOUCH VERIO METER READ INACCURATELY ERRATIC. THE PATIENT REPORTED BLOOD GLUCOSE RESULTS OF ¿10, 12.7, AND 13.4 MMOL/L¿ WITH THE SUBJECT METER, PERFORMED WITHIN 20 MINUTES OF EACH OTHER. THE PATIENT DID NOT EXPERIENCE ANY SYMPTOMS OR SEEK ANY MEDICAL ATTENTION BECAUSE OF THE REPORTED ISSUE. AS THE RESULTS FELL OUTSIDE EXPECTED VALUES FOR PRECISION TESTING, THIS COMPLAINT IS BEING REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
263710 OT VERIO METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL 3351640

Patients

Seq Age Sex Outcome Treatment
1 70 YR