XIA TITANIUM 4.5 EXTENDED CONNECTOR SMALL
Report
- Report Number
- 0009617544-2013-00216
- Event Type
- Injury
- Date Received
- June 11, 2013
- Date of Event
- May 13, 2013
- Report Date
- May 14, 2013
- Manufacturer
- STRYKER SPINE-FRANCE
- Product Code
- MNI
- PMA / PMN Number
- K052761
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
Narratives
ADDITIONAL INFORMATION HAS BEEN REQUESTED AND IF MADE AVAILABLE WILL BE REPORTED IN A SUPPLEMENTAL REPORT. METHOD, RESULT, AND CONCLUSION CODES WILL BE MADE AVAILABLE FOLLOWING AN ENGINEERING EVALUATION.
METHOD: DEVICE HISTORY REVIEW; COMPLAINT HISTORY REVIEW; RISK ASSESSMENT. RESULTS: THE CUSTOMER REPORTED EVENT OF XIA 4.5 EXTENDED CONNECTOR BREAKAGE, WAS CONFIRMED VIA X-RAYS TAKEN 5 MONTHS POST-OP AND PROVIDED FOR THE INVESTIGATION, AS THE PRODUCT WAS NOT RETURNED. A MANUFACTURING REVIEW DID NOT REVEAL ANY NONCONFORMITIES. THE CAUSE OF THE REPORTED FRACTURE OF THE XIA 4.5 EXTENDED CONNECTOR CANNOT BE CONCLUSIVELY DETERMINED. HOWEVER, REVIEW OF SIMILAR COMPLAINTS HAS REVEALED THAT THIS ISSUE IS OFTEN CAUSED BY FATIGUE BREAKAGE DUE TO THE STRUCTURE OF THE CONSTRUCT. CONCLUSION: THE CAUSE OF THE REPORTED FRACTURE OF THE XIA 4.5 EXTENDED CONNECTOR CANNOT BE CONCLUSIVELY DETERMINED. HOWEVER, REVIEW OF PREVIOUS SIMILAR COMPLAINTS HAS REVEALED THAT THIS ISSUE IS OFTEN CAUSED BY FATIGUE BREAKAGE DUE TO THE STRUCTURE OF THE CONSTRUCT.
IT WAS REPORTED THAT .. "ON (B)(6) 2010, THE PATIENT UNDERWENT THE SURGERY WITH THE SMALL XIA. ON (B)(6) 2013, THE SURGEON CONFIRMED X-RAY. AND HE FOUND THAT THE CONNECTOR BROKE. THEREFORE, THE SURGEON REMOVED THE BROKEN CONNECTOR AND THE PATIENT UNDERWENT THE REVISION SURGERY ON (B)(6) 2013. THE SURGEON REQUESTED THE INVESTIGATION OF THE BROKEN CONNECTOR..."
IT WAS REPORTED THAT ..."ON (B)(6) 2010, THE PATIENT UNDERWENT THE SURGERY WITH THE SMALL XIA. ON (B)(6) 2013, THE SURGEON CONFIRMED X-RAY. AND HE FOUND THAT THE CONNECTOR BROKE. THEREFORE, THE SURGEON REMOVED THE BROKEN CONNECTOR AND THE PATIENT UNDERWENT THE REVISION SURGERY ON (B)(6) 2013. THE SURGEON REQUESTED THE INVESTIGATION OF THE BROKEN CONNECTOR..."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 262017 | XIA TITANIUM 4.5 EXTENDED CONNECTOR SMALL | IMPLANT | MNI | STRYKER SPINE-FRANCE | 102152 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization| R |