FDA Adverse Event Injury Summary report: N

XIA TITANIUM 4.5 EXTENDED CONNECTOR SMALL

MDR report key: 3163009 · Received June 11, 2013

Report

Report Number
0009617544-2013-00216
Event Type
Injury
Date Received
June 11, 2013
Date of Event
May 13, 2013
Report Date
May 14, 2013
Manufacturer
STRYKER SPINE-FRANCE
Product Code
MNI
PMA / PMN Number
K052761
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION HAS BEEN REQUESTED AND IF MADE AVAILABLE WILL BE REPORTED IN A SUPPLEMENTAL REPORT. METHOD, RESULT, AND CONCLUSION CODES WILL BE MADE AVAILABLE FOLLOWING AN ENGINEERING EVALUATION.

Additional Manufacturer Narrative · 1

METHOD: DEVICE HISTORY REVIEW; COMPLAINT HISTORY REVIEW; RISK ASSESSMENT. RESULTS: THE CUSTOMER REPORTED EVENT OF XIA 4.5 EXTENDED CONNECTOR BREAKAGE, WAS CONFIRMED VIA X-RAYS TAKEN 5 MONTHS POST-OP AND PROVIDED FOR THE INVESTIGATION, AS THE PRODUCT WAS NOT RETURNED. A MANUFACTURING REVIEW DID NOT REVEAL ANY NONCONFORMITIES. THE CAUSE OF THE REPORTED FRACTURE OF THE XIA 4.5 EXTENDED CONNECTOR CANNOT BE CONCLUSIVELY DETERMINED. HOWEVER, REVIEW OF SIMILAR COMPLAINTS HAS REVEALED THAT THIS ISSUE IS OFTEN CAUSED BY FATIGUE BREAKAGE DUE TO THE STRUCTURE OF THE CONSTRUCT. CONCLUSION: THE CAUSE OF THE REPORTED FRACTURE OF THE XIA 4.5 EXTENDED CONNECTOR CANNOT BE CONCLUSIVELY DETERMINED. HOWEVER, REVIEW OF PREVIOUS SIMILAR COMPLAINTS HAS REVEALED THAT THIS ISSUE IS OFTEN CAUSED BY FATIGUE BREAKAGE DUE TO THE STRUCTURE OF THE CONSTRUCT.

Description of Event or Problem · 1

IT WAS REPORTED THAT .. "ON (B)(6) 2010, THE PATIENT UNDERWENT THE SURGERY WITH THE SMALL XIA. ON (B)(6) 2013, THE SURGEON CONFIRMED X-RAY. AND HE FOUND THAT THE CONNECTOR BROKE. THEREFORE, THE SURGEON REMOVED THE BROKEN CONNECTOR AND THE PATIENT UNDERWENT THE REVISION SURGERY ON (B)(6) 2013. THE SURGEON REQUESTED THE INVESTIGATION OF THE BROKEN CONNECTOR..."

Description of Event or Problem · 1

IT WAS REPORTED THAT ..."ON (B)(6) 2010, THE PATIENT UNDERWENT THE SURGERY WITH THE SMALL XIA. ON (B)(6) 2013, THE SURGEON CONFIRMED X-RAY. AND HE FOUND THAT THE CONNECTOR BROKE. THEREFORE, THE SURGEON REMOVED THE BROKEN CONNECTOR AND THE PATIENT UNDERWENT THE REVISION SURGERY ON (B)(6) 2013. THE SURGEON REQUESTED THE INVESTIGATION OF THE BROKEN CONNECTOR..."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
262017 XIA TITANIUM 4.5 EXTENDED CONNECTOR SMALL IMPLANT MNI STRYKER SPINE-FRANCE 102152

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R