ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-08094
- Event Type
- Injury
- Date Received
- June 11, 2013
- Date of Event
- May 10, 2013
- Report Date
- May 15, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN REQUESTED FOR RETURN TO ANIMAS AT THIS TIME. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
THE PATIENT CONTACTED ANIMAS ON (B)(6) 2013 REPORTING AN INCIDENT IN WHICH BLOOD GLUCOSE LEVELS DROPPED TO 20 MG/DL REQUIRING EMERGENCY MEDICAL ASSISTANCE AND HOSPITALIZATION. THE PATIENT REPORTED THAT DURING THE EVENT, THE PATIENT COULD FEEL BLOOD GLUCOSE LEVELS DROPPING RAPIDLY, FELT VERY DISORIENTED AND WAS ¿ACTING DRUNK.¿ THE PATIENT INDICATED THAT THEY DID NOT BELIEVE THAT THE LOW WAS CAUSED BY THE PUMP. THE PATIENT INDICATED HAVING A GREAT DEAL OF STRESS RECENTLY AND INDICATED RECENT WEIGHT LOSS. THE PATIENT REPORTED THAT THE HEALTH CARE PROVIDER ADVISED THE PATIENT NOT TO RESUME PUMP THERAPY UNTIL THE PATIENT CAN GAIN 10 LBS. THE PATIENT STATED THAT NO BOLUSES WERE DELIVERED LEADING UP TO THE LOW BLOOD GLUCOSE EVENT; THE TOTAL DAILY DOSE HISTORY CONFIRMED THAT ONLY BASAL RATE WAS DELIVERED. TROUBLESHOOTING OF THE EVENT INDICATED THAT THE PUMP APPEARED TO BE WORKING CORRECTLY. THIS REPORT IS MADE BASED ON THE ALLEGATION THAT THE PATIENT EXPERIENCED A SEVERE HYPOGLYCEMIC EVENT REQUIRING MEDICAL ATTENTION WHILE USING INSULIN PUMP THERAPY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 262707 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 58 YR | Hospitalization| L |