FDA Adverse Event Injury Summary report: N

QUADRA H FEMORAL STEM

MDR report key: 3162730 · Received June 5, 2013

Report

Report Number
3005180920-2013-00057
Event Type
Injury
Date Received
June 5, 2013
Manufacturer
MEDACTA INTERNATIONAL, SA
Product Code
LZO
PMA / PMN Number
K082792
Report Source
Manufacturer report

Narratives

Additional Manufacturer Narrative · 1

ON (B)(4) 2013, MEDACTA COLLECTED INFORMATION CONCERNING SIX CASES OF BONE FRACTURES IN WHICH ITS STEMS ARE INVOLVED. ALL OCCURRED TO THE SAME SURGEON DURING A CLINICAL STUDY. THE EVENTS OCCURRED DURING THE YEARS 2009, 2010, 2011, 2012 AND WERE NOT INITIALLY REPORTED TO MEDACTA AS COMPLAINTS. DOCUMENT REVIEW: WE ANALYZED OUR BATCH RECORDS OF EACH CASE, IN PARTICULAR FOR THIS ONE: QUADRA H FEMORAL STEM SIZE 7 - (B)(4)/LOT 080990 (50 STEMS PRODUCED). ALL PARAMETER WERE FOUND TO BE CONFORMING TO SPECIFICATIONS VALID AT THE TIME OF MANUFACTURING. THIRTY-EIGHT STEMS BELONGING TO THIS LOT HAVE BEEN ALREADY IMPLANTED AND NO INCIDENTS HAVE BEEN REPORTED UP TO NOW. ON THE BASIS OF THE DATA COLLECTED, THERE IS NO EVIDENCE THAT THE EVENT IS DEVICE RELATED, BUT IT CAN BE MORE RELATED TO SURGICAL MISTAKES. THE SIX CASES ARE REPORTED WITH THE MDR'S FROM 2013-00057 TO 2013-00062.

Description of Event or Problem · 1

REF IMP REPORT (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
249235 QUADRA H FEMORAL STEM FEMORAL STEM SIZE 7 STD CEMENTLESS LZO MEDACTA INTERNATIONAL, SA

Patients

Seq Age Sex Outcome Treatment
1