FDA Adverse Event Malfunction Summary report: N

PAINSMART IOD

MDR report key: 3162689 · Received May 13, 2013

Report

Report Number
1722139-2013-00740
Event Type
Malfunction
Date Received
May 13, 2013
Date of Event
January 7, 2008
Report Date
August 24, 2010
Manufacturer
MOOG DEVICE GROUP
Product Code
FRN
PMA / PMN Number
K981816
Removal / Correction Number
Z-1870-2011
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION FOUND THAT PUMP HAD EXPERIENCED ERROR CODE 45 IN PUMP'S HISTORY. NEW PUMP SOFTWARE WAS INSTALLED. REFERENCE RECALL NUMBER Z-1870-2011.

Description of Event or Problem · 1

INFO RECEIVED STATES THAT PUMP HAD EXPERIENCED ERROR CODE 45.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
211778 PAINSMART IOD FRN MOOG DEVICE GROUP PAINSMART IOD

Patients

Seq Age Sex Outcome Treatment
1