FDA Adverse Event
Malfunction
Summary report: N
PAINSMART IOD
MDR report key: 3162673
·
Received May 13, 2013
Report
- Report Number
- 1722139-2013-00742
- Event Type
- Malfunction
- Date Received
- May 13, 2013
- Date of Event
- May 16, 2010
- Report Date
- August 16, 2010
- Manufacturer
- MOOG DEVICE GROUP
- Product Code
- FRN
- PMA / PMN Number
- K981816
- Removal / Correction Number
- Z-1870-2011
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
INVESTIGATION FOUND THAT PUMP HAD EXPERIENCED ERROR CODE 45 IN PUMP'S HISTORY. NEW PUMP SOFTWARE WAS INSTALLED. REFERENCE RECALL NUMBER Z-1870-2011.
Description of Event or Problem · 1
INFO RECEIVED STATES THAT PUMP HAD EXPERIENCED ERROR CODE 45.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 211799 | PAINSMART IOD | FRN | MOOG DEVICE GROUP | PAINSMART IOD |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |