OLYMPUS THUNDERBEAT 5MM, 35CM, PISTOL GRIP
Report
- Report Number
- 8010047-2013-00028
- Event Type
- Malfunction
- Date Received
- February 16, 2013
- Date of Event
- January 18, 2013
- Report Date
- January 21, 2013
- Manufacturer
- OLYMPUS MEDICAL SYSTEMS CORPORATION
- Product Code
- GEI
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SN
- Reporter Occupation
- PHYSICIAN
Narratives
THE SUBJECT DEVICE WAS NOT RETURNED TO OLYMPUS MEDICAL SYSTEMS CORPORATION (OMSC) FOR EVALUATION. THE MANUFACTURING HISTORY WAS REVIEWED, WITH NO IRREGULARITIES NOTED. BASED ON THE REPORT BY OLYMPUS (B)(4) AND PAST SIMILAR CASES, IT IS POSSIBLE THAT SINCE THE DEVICE WAS KEPT ACTIVATED WHILE NOTHING WAS GRASPED IN THE GRASPING SECTION, THE PTFE PAD AND PROBE BECAME HOT, THE PTFE PAD WAS DEFORMED, AND PART OF THE DISTAL END OF THE PAD SEPARATED. THE INSTRUCTION MANUAL OF THUNDERBEAT MENTIONS WARNING AND CAUTION FOR POSSIBLE MISHANDLING MENTIONED ABOVE. DO NOT ACTIVATE OUTPUT FOR AN EXTENDED PERIOD WHILE THE GRASPING SECTION IS CLOSED WITHOUT CONTACTING TISSUE. OTHERWISE, A LOCAL INCREASE OF THE TEMPERATURE BETWEEN THE GRASPING SECTION DURING THE SEAL AND CUT MODE MAY RESULT IN PREMATURE WEAR, BREAKAGE AND/OR PROBE INSIDE THE BODY CAVITY. THIS REPORT IS BEING SUBMITTED AS A MEDICAL DEVICE REPORT IN AN ABUNDANCE OF CAUTION.
WHEN THE PHYSICIAN USED THE SUBJECT DEVICE IN A LAPAROSCOPY PROXIMAL GASTRECTOMY, THE PHYSICIAN NOTICED THE ABNORMALITY OF THE PTFE PAD OF THE SUBJECT DEVICE. THE PHYSICIAN REMOVED THE SUBJECT DEVICE FROM PATIENT AND CHECKED IT. THE PHYSICIAN FOUND THAT PTFE PAD PARTIALLY SEPARATED FROM THE GRASPING SECTION. THE PHYSICIAN ENSURED THAT NO PIECE OF THE PAD FELL OFF IN THE PATIENT'S BODY. THE INTENDED PROCEDURE WAS COMPLETED WITH A NEW THUNDERBEAT. THERE WAS NO PATIENT HARM REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 70224 | OLYMPUS THUNDERBEAT 5MM, 35CM, PISTOL GRIP | THUNDERBEAT HANDPIECE | GEI | OLYMPUS MEDICAL SYSTEMS CORPORATION | TB-0535PC | 23K |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |