FDA Adverse Event Malfunction Summary report: N

OLYMPUS THUNDERBEAT 5MM, 35CM, PISTOL GRIP

MDR report key: 3162597 · Received February 16, 2013

Report

Report Number
8010047-2013-00028
Event Type
Malfunction
Date Received
February 16, 2013
Date of Event
January 18, 2013
Report Date
January 21, 2013
Manufacturer
OLYMPUS MEDICAL SYSTEMS CORPORATION
Product Code
GEI
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SN
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE SUBJECT DEVICE WAS NOT RETURNED TO OLYMPUS MEDICAL SYSTEMS CORPORATION (OMSC) FOR EVALUATION. THE MANUFACTURING HISTORY WAS REVIEWED, WITH NO IRREGULARITIES NOTED. BASED ON THE REPORT BY OLYMPUS (B)(4) AND PAST SIMILAR CASES, IT IS POSSIBLE THAT SINCE THE DEVICE WAS KEPT ACTIVATED WHILE NOTHING WAS GRASPED IN THE GRASPING SECTION, THE PTFE PAD AND PROBE BECAME HOT, THE PTFE PAD WAS DEFORMED, AND PART OF THE DISTAL END OF THE PAD SEPARATED. THE INSTRUCTION MANUAL OF THUNDERBEAT MENTIONS WARNING AND CAUTION FOR POSSIBLE MISHANDLING MENTIONED ABOVE. DO NOT ACTIVATE OUTPUT FOR AN EXTENDED PERIOD WHILE THE GRASPING SECTION IS CLOSED WITHOUT CONTACTING TISSUE. OTHERWISE, A LOCAL INCREASE OF THE TEMPERATURE BETWEEN THE GRASPING SECTION DURING THE SEAL AND CUT MODE MAY RESULT IN PREMATURE WEAR, BREAKAGE AND/OR PROBE INSIDE THE BODY CAVITY. THIS REPORT IS BEING SUBMITTED AS A MEDICAL DEVICE REPORT IN AN ABUNDANCE OF CAUTION.

Description of Event or Problem · 1

WHEN THE PHYSICIAN USED THE SUBJECT DEVICE IN A LAPAROSCOPY PROXIMAL GASTRECTOMY, THE PHYSICIAN NOTICED THE ABNORMALITY OF THE PTFE PAD OF THE SUBJECT DEVICE. THE PHYSICIAN REMOVED THE SUBJECT DEVICE FROM PATIENT AND CHECKED IT. THE PHYSICIAN FOUND THAT PTFE PAD PARTIALLY SEPARATED FROM THE GRASPING SECTION. THE PHYSICIAN ENSURED THAT NO PIECE OF THE PAD FELL OFF IN THE PATIENT'S BODY. THE INTENDED PROCEDURE WAS COMPLETED WITH A NEW THUNDERBEAT. THERE WAS NO PATIENT HARM REPORTED.