ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-08076
- Event Type
- Malfunction
- Date Received
- June 11, 2013
- Report Date
- May 16, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- PATIENT
Narratives
FOLLOW-UP # 1 DATE OF SUBMISSION 06/11/2013 - DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: THE KEYPAD WAS FOUND TO BE FULLY INTACT; NO PEELING OR DAMAGE WAS OBSERVED. DURING TESTING, THE UP ARROW KEYPAD BUTTON WAS FOUND TO BE UNRESPONSIVE AND THE DOWN ARROW BUTTON WAS INTERMITTENTLY UNRESPONSIVE; THE OK AND CONTRAST BUTTONS RESPONDED APPROPRIATELY. THE KEYPAD WAS REMOVED AND THERE WAS EVIDENCE OF CONTAMINATION FOUND UNDER ALL KEY CONTACTS.
ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (TACTILE CHANGES/UNRESPONSIVE) ISSUE. IT WAS REPORTED THAT UP ARROW KEYPAD BUTTON WAS UNRESPONSIVE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 262131 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |