FDA Adverse Event Malfunction Summary report: N

HX2 TEMPERATURE MANAGEMENT SYSTEMM

MDR report key: 3162439 · Received June 6, 2013

Report

Report Number
1828100-2013-00598
Event Type
Malfunction
Date Received
June 6, 2013
Date of Event
May 16, 2013
Report Date
May 16, 2013
Manufacturer
TERUMO CARDIOVASCULAR SYSTEMS CORP.
Product Code
DWC
PMA / PMN Number
K071521
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE FIELD SERVICE REP (FSR) COULD NOT DUPLICATE THE REPORTED ISSUE. THE FSR REPLACED CHANNEL B MIX VALVE. WITH THE VALVE REPLACED AND PREVENTIVE MAINTENANCE COMPLETE, THE HEATER COOLER UNIT OPERATED TO MFR SPECIFICATIONS AND WAS RETURNED TO CLINICAL USE. IF ADDITIONAL INFORMATION BECOMES AVAILABLE ON THIS COMPLAINT THAT WOULD ALTER THE FACTS AND/OR CONCLUSION, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING SET-UP OF THE DEVICE FOR CARDIOPULMONARY BYPASS PROCEDURE, THE HEATER COOLER UNIT FAILED WITH AN AUDIBLE ALARM AND ERROR CODE OF "E0D" ON CHANNEL B. AS A RESULT, AN ALTERNATE DEVICE WAS EMPLOYED. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY, AND THERE WERE NO DELAYS, NO BLOOD LOSS OR NO ADVERSE CONSEQUENCES TO THE PT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
251744 HX2 TEMPERATURE MANAGEMENT SYSTEMM DWC TERUMO CARDIOVASCULAR SYSTEMS CORP. 809810

Patients

Seq Age Sex Outcome Treatment
1