FDA Adverse Event Malfunction Summary report: N

TERUMO ADVANCED PERFUSION SYSTEM 1

MDR report key: 3162434 · Received June 6, 2013

Report

Report Number
1828100-2013-00595
Event Type
Malfunction
Date Received
June 6, 2013
Date of Event
May 15, 2013
Report Date
May 15, 2013
Manufacturer
TERUMO CARDIOVASCULAR SYSTEMS CORP.
Product Code
DTQ
PMA / PMN Number
K022947
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE REPORTED COMPLAINT WAS CONFIRMED BY THE FIELD SERVICE REP (FSR). THE FSR REPLACED THE LID ASSEMBLY ON THE ROLLER PUMP. WITH THE LID ASSEMBLY REPLACED, THE UNIT OPERATED TO MFR SPECIFICATIONS AND WAS RETURNED TO CLINICAL USE. IF ADDITIONAL INFORMATION BECOMES AVAILABLE ON THIS COMPLAINT THAT WOULD ALTER THE FACTS AND/OR CONCLUSION, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY.

Description of Event or Problem · 1

THE FIELD SERVICE REP (FSR) REPORTED THAT DURING PREVENTATIVE MAINTENANCE (PM) OF THE DEVICE, THE LID SWITCH DETECTOR WAS FALLING OUT OF LID. SINCE THE EVENT OCCURRED DURING PREVENTATIVE MAINTENANCE, THERE WAS NO PT INVOLVEMENT.