ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-08075
- Event Type
- Malfunction
- Date Received
- June 11, 2013
- Report Date
- May 17, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
FOLLOW-UP # 1 DATE OF SUBMISSION 08/06/2013 - DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 07/15/2013 WITH THE FOLLOWING FINDINGS: DURING INVESTIGATION, THE PUMP HAD VISIBLE MOISTURE IN THE DISPLAY LENS. THE PUMP POWERED ON TO THE VERIFY SCREEN WITH AUDIBLE AND VIBRATORY SOUNDS. THE COMPLAINT OF A BLANK SCREEN WAS NOT DUPLICATED DURING INVESTIGATION. A LEAK TEST WAS PERFORMED AND A LEAK WAS FOUND DUE TO A CRACK BETWEEN UPPER RIGHT CORNER OF THE DISPLAY LENS AND THE CASE SEAL. THE PUMP COVER WAS REMOVED AND INTERNAL MOISTURE WAS PRESENT INSIDE THE PUMP. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (BLANK SCREEN W/MOISTURE) ISSUE. THE DISPLAY ON THE ANIMAS PUMP REPORTEDLY IS BLANK BUT STILL MAKES SOUNDS. THERE WAS NO TRAUMA ISSUE; HOWEVER, THERE WAS A MOISTURE ISSUE. THE CONTRAST WAS SET ACCORDINGLY BUT THE ISSUE IS NOT RESOLVED. THE PUMP IS BEING RETURNED FOR INVESTIGATION. THERE IS NO INDICATION THAT THE PRODUCT ISSUE CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ISSUE REMAINED UNRESOLVED AT THE TIME OF TROUBLESHOOTING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 263040 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 34 YR |