FDA Adverse Event Malfunction Summary report: N

ECHELON*FLEX60

MDR report key: 3162391 · Received June 11, 2013

Report

Report Number
3005075853-2013-02926
Event Type
Malfunction
Date Received
June 11, 2013
Date of Event
May 21, 2013
Report Date
May 22, 2013
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K081146
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). PREMATURE SLED MOVEMENT. ADDITIONAL INFORMATION WAS REQUESTED AND THE FOLLOWING WAS OBTAINED: ON WHICH FIRING DID THE EVENT OCCUR? UNKNOWN. IF THIS WAS A RELOAD FIRING, WERE THERE ANY OPENING ISSUES WITH PREVIOUS FIRINGS? UNKNOWN. AFTER USE, DID EACH OF THE TRIGGERS AND BUTTONS AUTOMATICALLY RETURN TO THEIR ORIGINAL (PRE-FIRED) POSITIONS, WITHOUT INTERVENTION? UNKNOWN. DID THE JAWS EVENTUALLY OPEN? UNKNOWN. WHAT COLOR CARTRIDGE WAS BEING USED? BLACK. WAS BUTTRESSING MATERIAL BEING UTILIZED? NO. IF YES, WHAT PRODUCT? THE DEVICE WAS RECEIVED FOR ANALYSIS WITH NO VISUAL NON-CONFORMANCES AND WITH AN ECR60T CARTRIDGE LOADED ON THE DEVICE. THE CARTRIDGE RELOAD WAS RECEIVED PARTIALLY FIRED 1/16. IT IS POSSIBLE THAT WHILE LOADING THE RELOAD, THE CARTRIDGE WAS PUSHED FARTHER BACK THAN THE CARTRIDGE ALIGNMENT STOP WINDOWS RESULTING IN THE KNIFE PUSHING THE ONE PIECE SLED FORWARD AND LOCKING THE CARTRIDGE. IT SHOULD BE NOTED THAT IN ORDER TO OPEN A DEVICE THAT HAS BEEN PARTIALLY FIRED OR FULLY FIRED A REVERSE STROKE NEEDS TO BE PERFORMED TRIGGER TO TRIGGER TO HANDLE IN ORDER TO RETURN THE KNIFE TO THE HOME POSITION (INDICATOR IN THE "0" POSITION) AND PRESS THE ANVIL RELEASE BUTTON TO OPEN. PLEASE NOTE THAT WHEN USING A TROCAR, THE INSTRUMENT JAWS MUST BE VISIBLE PAST THE TROCAR SLEEVE BEFORE OPENING THE JAWS. THE RETURNED DEVICE AND CARTRIDGE RELOAD WERE TESTED FOR FUNCTIONALITY IN THE ARTICULATED POSITION BY RESETTING AND RELOADING IT INTO THE DEVICE. THE DEVICE ACHIEVED ITS COMPLETE STROKE FIRING SEQUENCE WITHOUT ANY DIFFICULTIES. THE STAPLE LINE AND CUT LINE WERE COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORM SHAPE. THE EVENT COULD NOT BE CONFIRMED AS THE DEVICE CLOSED AND OPENED AS INTENDED. THE DEVICE HISTORY RECORDS WERE REVIEWED AND THE MANUFACTURING CRITERIA WAS MET PRIOR TO THE RELEASE OF THE BATCH.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC BOWEL RESECTION PROCEDURE, THE SURGEON COULD NOT OPEN THE DEVICE AFTER SHE PLACED IT IN THE TROCAR. CASE COMPLETED WITH ANOTHER DEVICE OF THE SAME PRODUCT CODE. THERE WERE NO PATIENT CONSEQUENCES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
262038 ECHELON*FLEX60 STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. NA UNK

Patients

Seq Age Sex Outcome Treatment
1