HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Report
- Report Number
- 1416980-2013-14970
- Event Type
- Malfunction
- Date Received
- June 11, 2013
- Date of Event
- May 18, 2013
- Report Date
- May 18, 2013
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE SAMPLE WAS NOT AVAILABLE FOR EVALUATION; THEREFORE, A DEVICE EVALUATION WILL NOT BE CONDUCTED. PROPER USE INSTRUCTIONS ARE PROVIDED IN THE HOMECHOICE AND HOMECHOICE PRO APD SYSTEMS PATIENT AT-HOME GUIDE WHICH IS SHIPPED WITH EVERY HOMECHOICE DEVICE. THE GUIDE INSTRUCTS THE USERS THAT CLAMPS ON ANY UNUSED SOLUTION LINES MUST REMAIN CLOSED WHEN PERFORMING PERITONEAL DIALYSIS THERAPY. A FORMAL REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. IF ADDITIONAL RELEVANT INFORMATION BECOMES AVAILABLE, A FOLLOW UP REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT A HOMECHOICE DEVICE EXPERIENCED A SYSTEM ERROR 2240 (AIR IN LINE). THIS OCCURRED DURING DWELL CYCLE FOUR OF FOUR IN PERITONEAL DIALYSIS THERAPY. THE PATIENT STATED THAT THE UNUSED FINAL LINE CLAMP WAS OPEN. THE TECHNICAL SERVICES REPRESENTATIVE (TSR) ASSISTED THE PATIENT IN CLEARING THE ALARM AND ADVISED THEM TO START THERAPY OVER WITH NEW SUPPLIES. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION REPORTED. ADDITIONAL INFORMATION WAS REQUESTED AND IS NOT AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 262601 | HOMECHOICE AUTOMATED PD SET WITH CASSETTE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 83 YR | HOMECHOICE |