FDA Adverse Event Malfunction Summary report: N

XIENCE PRIME EVEROLIMUS ELUTING CORONARY STENT SYSTEM

MDR report key: 3162323 · Received June 11, 2013

Report

Report Number
2024168-2013-03639
Event Type
Malfunction
Date Received
June 11, 2013
Date of Event
May 16, 2013
Report Date
May 17, 2013
Manufacturer
AV-TEMECULA-CT
Product Code
NIQ
PMA / PMN Number
P110019
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR EVALUATION. DIFFICULTY/RESISTANCE REMOVING THE DEVICE POST-DEPLOYMENT COULD NOT BE REPLICATED IN A TESTING ENVIRONMENT AS IT WAS BASED ON OPERATIONAL CIRCUMSTANCES. BASED ON VISUAL ANALYSIS OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. REVIEW OF THE COMPLAINT HISTORY OF THE REPORTED LOT DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT YET COMPLETE. A FOLLOW UP REPORT WILL BE SUBMITTED WITH ALL RELEVANT INFORMATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE TARGET LESION WAS A CHRONIC TOTAL OCCLUSION (CTO) IN THE RIGHT CORONARY ARTERY (RCA) WITH HEAVY CALCIFICATION. PRE-DILATATION WAS PERFORMED WITH A MINI TREK 2.0 X 20 MM BALLOON AT 12 ATMOSPHERES (ATM). THE XIENCE PRIME 4.0 X 23 MM WAS IMPLANTED WITHOUT ISSUE. HOWEVER, DURING RETRACTION OF THE STENT DELIVERY SYSTEM (SDS) BALLOON, THE SDS BALLOON WAS STICKING TO THE IMPLANTED STENT. THE SDS WAS ABLE TO BE RETRACTED BY CAREFULLY MOVING THE BALLOON. THE XIENCE PRIME STENT WAS CONFIRMED TO BE IMPLANTED SUCCESSFULLY AND THE INTERVENTION WAS COMPLETED WITH A GOOD RESULT. NO ADVERSE PATIENT EFFECTS WERE REPORTED. THERE WAS NO CLINICALLY SIGNIFICANT DELAY OF PROCEDURE REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
262353 XIENCE PRIME EVEROLIMUS ELUTING CORONARY STENT SYSTEM DRUG ELUTING CORONARY STENT SYSTEM NIQ AV-TEMECULA-CT 2030841

Patients

Seq Age Sex Outcome Treatment
1 GUIDE CATH: 8F