FDA Adverse Event
Injury
Summary report: N
SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
MDR report key: 3162285
·
Received June 11, 2013
Report
- Report Number
- 1416980-2013-14951
- Event Type
- Injury
- Date Received
- June 11, 2013
- Date of Event
- May 10, 2013
- Report Date
- May 18, 2013
- Manufacturer
- BAXTER HEALTHCARE
- Product Code
- KDJ
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE CAUSE OF THIS PERITONITIS WAS USE ERROR. PER BAXTER LABELING, USERS ARE INSTRUCTED TO USE ASEPTIC TECHNIQUE WHEN PERFORMING PERITONEAL DIALYSIS THERAPY. A FORMAL REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. IF THERE IS ANY FURTHER RELEVANT INFORMATION FROM THAT REVIEW, A SUPPLEMENTAL MEDWATCH WILL BE FILED.
Description of Event or Problem · 1
THIS IS A REPORT OF A HOME PATIENT (HP) WHO WAS HOSPITALIZED AND DIAGNOSED WITH PERITONITIS. THE CAUSE OF THE PERITONITIS WAS A BREAK IN ASEPTIC TECHNIQUE, FURTHER DESCRIBED AS THE ATTENDANT PERFORMED THERAPY WITHOUT PROPER TRAINING. THE HP WAS TREATED WITH INJECTIONS OF VENCOGEN-IP 1GM, REFLIN 1GM AND FORTUM 1GM (FREQUENCIES WERE NOT REPORTED). THE OUTCOME FOR THIS EVENT IS UNKNOWN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 262677 | SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE | KDJ | BAXTER HEALTHCARE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 72 YR | Hospitalization| R | DIANEAL 2.5% ULTRABAGS |