FDA Adverse Event Malfunction Summary report: N

STENOSCOP

MDR report key: 3162211 · Received June 3, 2013

Report

Report Number
1720753-2013-06652
Event Type
Malfunction
Date Received
June 3, 2013
Date of Event
May 21, 2013
Report Date
June 3, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE LEFT MONITOR BNC CABLE WAS REPLACED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE IMAGE WAS LOST WHEN THE LEFT MONITOR WAS MOVED. THIS RESULTED IN A LOSS OF LIVE IMAGE. THERE IS NO REPORT OF PT INJURY ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
242495 STENOSCOP FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) STENOSCOP

Patients

Seq Age Sex Outcome Treatment
1