FDA Adverse Event
Malfunction
Summary report: N
ONETOUCHPING GLUCOSEMGMTSYSTEM
MDR report key: 3161968
·
Received June 11, 2013
Report
- Report Number
- 2531779-2013-08062
- Event Type
- Malfunction
- Date Received
- June 11, 2013
- Report Date
- May 16, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
FOLLOW-UP # 1 DATE OF SUBMISSION (B)(4) 2013 - DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: DURING INVESTIGATION, THE DISPLAY SCREEN WAS OBSERVED TO BE CRACKED. THE PUMP POWERED ON WITH THE DISPLAY REMAINING BLANK. WHEN A TEST DISPLAY SCREEN WAS USED FOR FURTHER EVALUATION, THE DISPLAY FUNCTIONED PROPERLY.
Description of Event or Problem · 1
ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (DAMAGED) ISSUE. REPORTEDLY, THE DISPLAY SCREEN WAS CRACKED AND WAS DIFFICULT TO READ. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 263169 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 13 YR |