Description of Event or Problem · 1
ONGOING PROBLEMS WITH ATHENA NEONATAL 9050 MONITOR. DEVICE 1 11/00 MONITOR SHUT DOWN FOR NO APPARENT REASON. DETERMINED THAT MONITOR HAD A "BAD POWER SUPPLY", REPLACED AND POWER WAS RESTORED. DEVICE 2: 11/00 MONITOR SHUT DOWN FOR NO APPARENT REASON. POWER DID NOT RETURN TO MONITOR. CONCLUDED MONITOR HAS A BAD HV DEFLECTION BOARD. DEVICE 3: 11/00, MONITOR SHUT DOWN - INTERNITTENT POWER SUPPLY. DEVICE 4: 11/00, MONITOR SHUT DOWN - INTERMITTEND POWER SUPPLY. DEVICE 5: 11/00, MONITOR SHUT DOWN - HV DEFLECTION BOARD. DEVICE 6: 11/00, MONITOR SHUT DOWN - HV DEFLECTION BOARD. DEVICE 7: 12/00, MONITOR SHUT DOWN - HV DEFLECTION BOARD. DEVICE 8: 12/01, MONITOR HAD ALARM LIMITS SET. EKG ALARM WENT OFF MANY TIMES WHEN NO ALARM CONDITION EXISTED. CONCERNS: CARDIOPULMONARY NEONATAL MONITORS FAILING WITHOUT WARNING, PUTTING THE SAFETY OF THE NEONATES IN JEOPARDY. 8 OUT OF 25 MONITORS HAVE FAILED OVER THE PAST 3 MONTHS. ORIGINAL MFR SOLD THIS LINE OF MONITORS. PARTS ARE DIFFICULT TO OBTAIN. REPLACEMENT PARTS COME FROM OVERSEAS AND USUALLY TAKES 4-5 MONTHS TO RECEIVE THEM. SVC BULLETIN, DATED 10/30/96, WHICH HOSP RECENTLY BECAME AWARE OF, SUGGESTS A FAULTY CAPACITOR (C206) WILL RESULT IN A BLANK SCREEN AFTER THE MONITOR IS SWITCHED ON. BULLETIN IDENTIFIED SERIAL NUMBERS INVOLVED, BUT SOME OF HOSP'S ARE OUTSIDE THOSE NUMBERS. EXTEND OF PROBLEM IS UNKNOWN. NOT CLEAR WHO (MFR/DISTRIBUTOR) OWNS THE PROBLEMS. UNCLEAR IF THERE IS A SECOND PROBLEM. HOSP KNOWS THE CAPACITOR IS CAUSING SOME OF THE PROBLEMS.