FDA Adverse Event
Malfunction
Summary report: N
9900
MDR report key: 3161698
·
Received June 3, 2013
Report
- Report Number
- 1720753-2013-06720
- Event Type
- Malfunction
- Date Received
- June 3, 2013
- Date of Event
- May 23, 2013
- Report Date
- June 4, 2013
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE GENERAL PURPOSE OPERATING SYSTEM, REAL TIME OPERATING SYSTEM AND CMOS BATTERIES WERE REPLACED, AND THE CMOS SETTINGS WERE RELOADED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT WHEN THEY TURNED THE MACHINE ON, THEY WERE HAVING TROUBLE WITH IT NOT BOOTING UP ALL THE WAY. THERE IS NO REPORT OF PATIENT INJURY ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 242593 | 9900 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |